background image

INSTRUCTION MANUAL

HI98318

Waterproof EC/TDS Tester

Warranty

These meters are warranted for a period of one year against defects in workmanship 
and materials when used for their intended purpose and maintained according to 
instructions. Probes are warranted for a period of six months. This warranty is limited 
to repair or replacement free of charge. Damage due to accidents, misuse, tampering 
or lack of prescribed maintenance is not covered. If service is required, contact your 
local Hanna Instruments Office. If under warranty, report the model number, date 
of purchase, serial number and the nature of the problem. If the repair is not covered 
by the warranty, you will be notified of the charges incurred. If the instrument is to 
be returned to Hanna Instruments, first obtain a Returned Goods Authorization 
(RGA) number from the Technical Service department and then send it with shipping 
costs prepaid. When shipping any instrument, make sure it is properly packaged for 
complete protection.

Solutions

Code

Description

HI5036-023

Quick Cal pH and EC one-point 

calibration solution,  

230 mL bottle

HI50036P

Quick Cal pH and EC one-point

calibration solution,  

20 mL sachets (25 pcs.)

HI70031P

1413 μS/cm calibration solution,

 

20 mL sachets (25 pcs.)

HI7031L

1413 μS/cm calibration solution,  

500 mL bottle

HI8031L

1413 μS/cm calibration solution, FDA approved,

500 mL bottle

Accessories

 

 

IST98318   09/18

To obtain the highest accuracy for measurements it is important to follow these tips:

•  Calibration is only as good as the solution being used. The calibration solution 

values change over time. Fresh solution should be used for each calibration.

•  The probe should be rinsed with purified water each time before placing in 

calibration solution or sample to be tested.

Care and Maintenance

Battery Replacement

To change the CR2032 Li-ion battery, turn the battery cover located on the back of 

the meter counterclockwise to unlock. Remove cover and replace with new battery 

+ side facing up.

Note: Batteries should only be replaced in a safe area using the battery 

type specified in this instruction manual. Old batteries should be disposed in 

accordance with local regulations.

The meter features a low battery indicator. When the battery is running low (under 

10%), the battery indicator will blink on the LCD. When the battery is discharged  

“dEAd bAtt”

 will be displayed on the LCD for 2 seconds and the meter will turn off.

Error messages

During user calibration, if the reading is out of the accepted range, the meter will 

display “---- Err“.i.e. 3000 ppm blinking means the limit of the measured value is 

out of range.
If the measured temperature is higher than 50.0 ºC or lower than 0.0 ºC, the 50.0 ºC or 

0.0 ºC temperature value will blink on the LCD in measure mode.

2. Place the probe in calibration solution. The probe tip should be centered in the 

solution and submersed 1.18” (3 cm). The meter will automatically recognize the 

solution.

3. If the solution is not recognized or is out of the accepted range “

---- 

Err“ is 

displayed.

4. If the buffer is recognized “REC“ is displayed until the reading is stable and the 

calibration is accepted.

• After acceptance, the “Stor“ message is displayed and meter returns to 

measurement mode.

Calibration Procedure

EC Measurement and Calibration

For better accuracy, frequent calibration of the meter is recommended. In addition, 

the instrument must be recalibrated:

• When high accuracy is required.
• At least once a month.

Measurement

Press the ON/OFF button to turn the meter ON. Place the probe to the desired solution. 

The conductivity value, automatically compensated for temperature, will be shown 

on the LCD.

The measured temperature will be shown on the secondary LCD.

Note: Before taking any measurement make sure the meter has been calibrated.

• Press CAL button to enter calibration mode from measurement mode. 
• The meter will enter the calibration mode, displaying “mS/cm 1.41 USE” with 

CAL tag blinking.

1. Pour 2” (5 cm) of standard solution into a container.

• After entering in the calibration mode and before the point is accepted, it is 

possible to quit the procedure and return to the last calibration data by pressing 

the CAL button. The LCD displays ”----ESC” for 1 second and the meter returns to 

measurement mode.

Quick Calibration

Select calibration type “CAL quic “ from meter setup.

• Enter calibration mode.
• Immerse the probe in the 

HI50036

 calibration solution.

• When the standard values is recognized and stability is reached the meter 

automatically accepts the calibration.

• The LCD will display “Stor“ for 1 second and return to normal measurement mode.
• If the standard is not recognized or the slope is out of accepted range “----Err‘ is 

displayed. Change the calibration solution, clean the electrode or press any key 

to exit calibration.

• When the calibration procedure is completed, the “CAL“ tag is turned on.

Clear Calibration

Place meter in calibration mode. Press ON/OFF and “CLR” is displayed. The meter 

will now be at default calibration. 