background image

Instruction Manual

(HI 981404, HI 981405)

 pH & TDS/EC Monitors

with Visual Alarm

http://www.hannacan.com

These Instruments are in

Compliance with the CE Directives

Hanna Instruments reserves the right to modify the design, construction

and appearance of its products without advance notice.

WARRANTY

WARRANTY

HI 981404 and HI 981405 are warranted for two years 

against

defects in workmanship and materials when used for their intended

purpose and maintained according to instructions.

 The probes are

warranted for a period of one year.

 This warranty is limited to

repair or replacement free of charge.

Damages due to accident, misuse, tampering or lack of prescribed

maintenance are not covered.

If service is required, contact the dealer from whom you purchased the

instrument. If under warranty, report the model number, date of pur-

chase, serial number and the nature of the failure. If the repair is not

covered by the warranty, you will be notified of the charges incurred. If

the instrument is to be returned to Hanna Instruments, first obtain a

Returned Goods Authorization Number from the Customer Service de-

partment and then send it with shipment costs prepaid. When shipping

any instrument, make sure it is properly packaged for complete protec-

tion.

To validate your warranty, fill out and return the enclosed warranty card

within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited

without the written consent of the copyright owner, Hanna Instru-

ments Inc.

Dear Customer,
Thank you for choosing a Hanna product. This manual will

provide you with the necessary information for the correct

operation of the meter. Please read it carefully before using

the meter. If you need additional technical information, do

not hesitate to e-mail us at [email protected].
These instruments are in compliance with the   directives

EN 50081-1 and EN 50082-1.

Remove the instrument from the packing material and

examine it carefully. If any damage has occurred during

shipment, immediately notify your Dealer or the nearest

Hanna Customer Service Center.
The meter is supplied with:

• HI 1286 pH electrode

• HI 1283 grounding probe

• HI 7634 TDS probe (HI 981404)

• HI 7632 EC probe (HI 981405)

• Calibration screwdriver

• 12 VDC power adapter

Note:

Conserve all packing material until the instrument

has been observed to function correctly. Any defective

item must be returned in its original packing.

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

HI 981404 

(pH and TDS) and 

HI 981405 

(pH and EC)  are

specially designed to meet the needs of simple continuous

monitoring.
The housing has an IP54 rating.
You can simply install the meter above the sample to be

tested for continuous measurement.
The HI 1286  gel-filled pH electrode is replaceable and the

BNC connector is protected behind a waterproof sheath. The

unique design of the electrode provides longer life in aggres-

sive solutions.
You can select your own setpoint and be informed when a

dosage is required by a flashing LED alarm.
Measurements are accurate and the meter can be cali-

brated at one or two points for pH and at a single point for

TDS and EC.
You no longer need to worry about battery changes either:

the unit runs without interruption on 12 VDC supplied

adapter.

GENERAL DESCRIPTION

GENERAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

1. Liquid Crystal Displays

2. Alarm LED for pH and PPM / EC

3. pH 7.0 calibration trimmer

4. pH 4.0/10.0 calibration trimmer

5. EC / TDS calibration trimmer

6. BNC connector

7. Protective sheath

8. HI 1286

9. HI 1283 - Grounding probe

10. HI 7632 / HI 7634

 11. Power supply connector

 12. 12 VDC power adapter

5.8

pH

1190

HI 981404 / HI 981405

HI 981404 / HI 981405

Range

0.0 to 14.0 pH

Resolution

0.1 pH

Accuracy (@ 25°C/77°F)

±0.2 pH

Setpoint

3.0 to 7.0 pH

Alarm

     LED blinks when pH is higher than setpoint

Calibration

          Manual at 1 or 2 points

Probes    

HI 1286 replaceable pH electrode and HI

1283 grounding probe (included)

Power supply

External 12 VDC (adaptor included)

Dimensions

165 x 110 x 35 mm (6.5 x 4.3 x 1.4’’)

Weight

300 g (10.6 oz.)

HI 981404 only

HI 981404 only

Range              

0 to 1990 mg/L (ppm)

Resolution

10 mg/L (ppm)

Accuracy (@ 25°C/77°F)

±2% f.s.

TDS Factor

0.7 ppm = 1 µS/cm

Temp. Compensation

     Automatic 5 to 50°C

                  (41 to 122°F, TDS only)

Setpoint

500 to 1600 mg/L (ppm)

Alarm      

LED blinks when TDS is lower than setpoint

Calibration

Manual at 1 point

Probe

HI 7634 TDS probe (included)

HI 981405 only

HI 981405 only

Range             

 0.00 to 9.99 mS/cm (EC)

Resolution

          0.01 mS/cm (EC)

Accuracy (@ 25°C/77°F)

±2% f.s.

Temp. Compensation

     Automatic 5 to 50

°

C

                  (41 to 122

°

F, EC only)

Setpoint

1.00 to 3.50 mS/cm (EC)

Alarm

     LED blinks when EC is lower than setpoint

Calibration

               Manual at 1 point

Probe

HI 7632 EC probe (included)

SPECIFICATIONS

SPECIFICATIONS

5.8

pH

1190

1

2

3

4

5

6

7

8

9

1 1

1 0

1 2

Отзывы: