Operational Guide
45
If the reading is out of range, the full-scale value
(400.0%) will be displayed blinking.
If this occurs during a log, a “!“ will be placed next to
the measurement unit. Data found in the CSV files
with a “!“ should not be considered reliable.
Natural Sea Water Scale
The Natural Sea Water Scale extends from 0 - 80.0 g/L. It determines salinity
based upon a conductivity ratio of sample to “standard seawater” at 15 °C.
Where R
15
is the conductivity ratio and salinity is defined by the following
equation:
S = - 0.08996 + 28.2929729R
15
+ 12.80832R
15
2
- 10.67869R
15
3
+ 5.98624R
15
4
-1.32311R
15
5
Note
: The formula can be applied for temperatures
between 10 ºC and 31 ºC.
If the reading is out of the measurement range, the display will flash the highest
value possible and a warning message will be displayed. If this occurs during a
log a “!” will be placed next to the measurement unit. Data found in the CSV file
with a ”!” should not be considered reliable.
Salinity
Measurements
www.
.com