
Operational Guide
30
Note
: If a custom buffer was used in calibration, the “C1” and “C2” tags will be
displayed. If only the one custom buffer is used in calibration, the tag will be “C1”
and the value will be displayed.
If disabled, “EXPIRATION WARNING DISABLED” is
displayed.
Or if enabled the number of days until the calibration
alarm “CAL DUE” will be displayed. (i.e. “CAL EXPIRES
IN 2 DAYS”)
The number of days since the calibration expired. (i.e.
“CAL EXPIRED 2 DAYS AGO”).
The probe serial number together with the current
reading.
If a buffer is not from the last calibration, the buffer
tag will be displayed blinking.
In Standard Mode, Condition and Response gauges are visible on the day of
calibration (See Electrode Condition And Electrode Response Time page 24).
If configured in SETUP, a countdown message displays the number of days
remaining until a new Calibration is due.
If using a HI 11311 or HI 12301 sensor, the glass
impedance is continuously monitored, updated and
reported here and the Junction Condition Gauge is
fully visible.
pH GLP
Information
www.
.com
1.800.561.8187