32
Symptoms
Problems
Solution
Readings fluctuate up and
down (noise).
DO: DO probe electrolyte
contains entrapped gas.
Remove cap. Refill, tap and
reinstall.
The display shows DO
reading blinking.
Out of range in DO scale.
Verify area of cathode is free of
bubbles inside cap. Verify solution
movement past membrane. Remove
cap, inspect and clean if necessary.
Install new cap, fresh electrolyte
with no bubbles, permit longer
polarization. Stir or increase flow
rate.
The meter does not
measure temperature. “‑‑
‑‑” is displayed on second
LCD line.
Broken temperature sensor. Replace the probe.
At startup the meter
displays all LCD tags
permanently.
One of the keys is stuck.
Check the keyboard or contact your
local Hanna Instruments Office.
CAL “Prod” message at
startup.
Instrument was not factory
calibrated or lost factory
calibration.
Contact your local Hanna
Instruments Office for help.
TROUBLESHOOTING GUIDE
MAINTENANCE