Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI772

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 25 tests
• 1 mL syringe with tip
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0.0”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you use 
the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must 

be replaced. Once this indication is displayed, normal operation of 
the instrument will be interrupted. Change the battery and restart 
the meter.

 

For more details about spare parts and accessories see “Accessories”.

HI772

  

Marine Alkalinity

INSTRUCTION MANUAL

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument. 

For more information about Hanna Instruments and our products, visit  
www.hannainst.com.

For technical support, contact your local Hanna Instruments Office or e-mail us at  
[email protected]

Preliminary Examination

Specifications

Range

0.0 to 20.0 dKH

Resolution

0.1 dKH

Accuracy

±0.3 dKH  ±5% of reading @ 25 ˚C / 77 ˚F

Light Source

light emitting diode @ 610 nm

Light Detector

silicon photocell

Method

Colorimetric method. The reaction causes a distinctive range of 

colors from yellow to green to greenish blue to develop

Environment

0 to 50 °C (32 to 122 °F); RH max 95% non-condensing

Battery Type

1.5V AAA (1 pc.)

Auto-off

after 10 minutes of non-use

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

64 g (2.3 oz)

Functional Description

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Errors and Warnings