SETUP
7. SETUP
• Short press MENU key to enter in Setup and to move to next item in menu.
• Long press MENU key to exit.
• Press arrow keys to change the values.
• Short press MENU key to automatically save modified values.
Table below presents an overview of
BL100
menu with ranges and factory set defaults.
Parameter
(Scrolled message)
Range / Option
Default settings
Control
Auto or oFF
Auto
Control type
on/oFF
Proportional
on/oFF
Hysteresis
0.1 to 1.00 pH
–
0.50
Band
–
0.1 to 2.00 pH
1.0
Set point
6.00 to 8.00
7.20
Control mode
Hi or Lo
Hi
Start up delay (seconds)
0 to 600
60
Overtime alarm (minutes)
oFF, 0 to 180
30
Flow rate (L/H)
Flow rate (G/H)
0.5 to 3.5
0.13 to 0.92
1.0
0.26
Level alarm
diS or En
diS
High alarm
diS or En
En
High alarm value (pH)
0 to 14.00 pH*
8.00 pH
Low alarm
diS or En
diS
Low alarm value (pH)
0 to 13.90 pH*
6.00 pH
Temperature unit
°C or °F
°C
Flow rate measurement unit
L / H or G / H
L / H
Language
En (English), ES (Spanish),
Fr (French), Pt (Portuguese),
nL (Dutch)
En (English)
*Available range changes, based upon other settings.
High alarm must be set higher than low alarm. If low alarm was set 7 pH, then high alarm range
is 7.1 to 14 pH.