background image

 

 

T04B

LOW-SPEED CENTRIFUGE

USER MANUAL

Содержание T04B

Страница 1: ...T04B LOW SPEED CENTRIFUGE USER MANUAL...

Страница 2: ...Product name Centrifuge Model name T04B Manufacturer Hanil Scientific Inc Address 16 Arayukro Gimpo 10136 Korea Research Use Only LOW SPEED CENTRIFUGE T04B www ihanil com UM T04B E Rev 1 20210406...

Страница 3: ...lse Short Spin 15 4 6 Start Stop 16 4 7 Change setting value during operation 17 4 8 Key Lock setting 17 4 9 Save and recall program 17 4 10 Sound level setting 18 4 11 Set the number of end tones 19...

Страница 4: ...s manual describes the performance usage and handling precautions of the purchased product Please read carefully before using the product Rotor insertion tightning caution indication warning Bucket mo...

Страница 5: ...ubstances that can generate volatile or explosive vapors cannot be centrifuged 8 When the rotor comes into contact with cleaning solutions such as strong acids or strong bases or cesium silver salt it...

Страница 6: ...techsupport ihanil com 2 1 Appearance 2 2 Components User manual AC Power Cord Rotor Locking Tool Emergency Lid Open Tool 2 Product composition and information Grease Lubricant Rotor optional Manual...

Страница 7: ...USER MANUAL T04B A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com 6...

Страница 8: ...USER MANUAL T04B 7 ihanil com A S Inquiry 02 3452 8966 techsupport ihanil com...

Страница 9: ...f the main body and connect the power plug to the outlet Please check the rated voltage to be used 2 Press the power switch button I O located on the right side of the main unit in the ON direction I...

Страница 10: ...ing the rotor to the central shaft in the chamber fix it using the provided rotor locking tool Rotor mounting clockwise rotation Rotor removal counterclockwise rotation Hold the rotor with one hand an...

Страница 11: ...or centrifuge recommended by Hanil Scientific Inc and do not use it above the allowable standard after checking the maximum RCF value for each tube Samples are accurately measured in the same amount d...

Страница 12: ...orrectly 1 If you press the Program button for 2 seconds the Rotor information window appears While the Rotor Information window is displayed enter the Rotor information using the number buttons 2 Sam...

Страница 13: ...tion is possible Control interval 1 Min 1 Sec START STOP It is used to start and stop centrifugation ACC DEC The acceleration speed in the start and stop section can be set in 9 steps and the decelera...

Страница 14: ...inutes 59 seconds or continuous operation 00 minutes 00 seconds setting is possible In addition AT SET SPEED time mode time counting after reaching the set speed is supported for accurate time managem...

Страница 15: ...cond input mode switching 1 Press TIME once 2 To enter seconds enter the setting value using the numeric keypad and press Enter Press Enter Switch to the minute input mode Changes in 1 second incremen...

Страница 16: ...ue use the numeric keypad to enter the setting value and press Enter ACC is set in 1 9 steps ACC 9 fastest acceleration section DEC is set in 0 9 steps DEC 0 natural deceleration Press Enter final sto...

Страница 17: ...lid is closed When Enter is input during operation it indicates the set value just before the start button is pressed 4 6 2 Stop 1 If you want to end the operation press the Start Stop button During d...

Страница 18: ...during operation The changed time input value does not reflect the initial time setting value 4 8 Key lock setting 4 7 Change setting value during operation 4 9 Save and recall program Program storag...

Страница 19: ...e set values speed time acc dec etc of the called program are displayed on the display screen If you do not enter the setting value for 15 seconds the setting mode is canceled 4 10 Sound setting The l...

Страница 20: ...t be performed after the machine has completely stopped rotating If this is not followed it may damage the sample and the user so be careful Do not close the lid immediately after emergency opening wa...

Страница 21: ...1 If an acid basic solution or solution spills from the tube immediately wipe it off with a soft cloth dampened with warm water and store in a dry place 2 The tube hole of the Fixed Angle rotor or th...

Страница 22: ...e in Emergency for sample protection Check if there is any foreign substance on the lid clasp if there is remove the foreign substance and close the lid If the installation position of the main body i...

Страница 23: ...again to check the operation status again Lid error Occurs when the Lid Sensor does not operate normally Error 20 27 1 Check that the weight of the sample inserted in the rotor is the same and that i...

Страница 24: ...ubecapacity mL 14mL 15 15mL conical 25mL conical 25mL conical 30 50 50mL conical 50mL conical Skirt TubeDimension xL mm 15 7x96 16x120 17x120 28 8x83 28 8x78 5 25 7x101 4 29x108 29 5x118 29 5x118 Adap...

Страница 25: ...116 3 Outer 132 3 Inner 136 8 Outer152 8 Inner 105 3 Outer121 3 Inner 105 3 Outer121 3 Inner 126 5 Outer142 5 Inner 138 Outer155 Inner 138 Outer155 Inner 138 Outer155 Max RCF g force Inner2 080 Outer2...

Страница 26: ...al Max heighttube fit mm 115 115 75 75 115 115 120 120 120 Max radius mm 150 1 150 1 110 1 110 1 150 1 150 1 152 1 155 1 155 1 Max RCF g force 2 685 2 685 1 969 1 969 2 685 2 685 2 721 2 774 2 774 Tub...

Страница 27: ...tomtype Conical Conical Conical Round Round Max radius mm 152 5 118 5 118 5 143 9 144 1 155 1 Max RCF g force 2 728 2 120 2 120 2 574 2 578 2 774 Tube Tubecapacity mL 5mLconical 5mLconical 14mL 15mLco...

Страница 28: ...onical Round Max heighttubefit mm 88 115 75 75 103 5 115 115 115 125 center Max radius mm 125 156 6 112 112 140 5 152 152 152 152 Max RCF g force 2 236 2 719 2 003 2 003 2 513 2 719 2 719 2 719 2 719...

Страница 29: ...TR30 50c TR50 50c Adaptorholedimension xL mm 17x105 27 1x14 1 27 1x14 1 26x83 8 27 9x11 Adaptorholebottomtype Conical Conical Conical Round Round Max radius mm 150 9 119 9 119 9 145 3 145 5 150 9 Max...

Страница 30: ...otor S mw 2 Tube Tubecapacity mL 2 0 4 mL VT 4 7 mL VT 8 10 mL VT 15 mL glass 15 mL open top TubeDimension xL mm 13 x 75 13 x 100 16 x 100 16 x 100 16 x 114 Adapter None None None CatNo TR3 15 TR5 15...

Страница 31: ...A S Inquiry 02 3452 8966 techsupport ihanil com USER MANUAL T04B ihanil com 30 MEMO...

Страница 32: ......

Отзывы: