TECHNICAL SPECIFICATIONS
Technical Data - OmniSWD
®
GENERAL:
Power Supply:
120VAC 50/60Hz
Power consumption:
250W max
Fuses:
Externally Replaceable
2 x T 2A 20mm
Size:
Approx. H: 870 mm (34.25 in) / D: 410 mm (16 in) / W: 410 mm (16 in)
Weight:
39 kg (86 lbs.)
FDA Classification:
Class II: Type BF
Frequency:
27.12 MHz + 20 KHz
Output Power:
200W Peak Power (+20%) into 50 Ohm load pulsed mode
Output Modes:
Pulsed output waveforms are available as selected by the user interface. In pulsed mode the output is square
wave modulated.
Pulse modes presets:
Protocol 1ΔT
- Timer is preset to 15 minutes, pulse rate is set 1
T to 800 PPS and pulse duration to 100 µSec.
Protocol 2ΔT
- Timer is preset to 15 minutes, pulse rate is set 2
T to 800 PPS and pulse duration to 200 µSec.
Protocol 4ΔT
- Timer is preset to 15 minutes, pulse rate is set 4
T to 800 PPS and pulse duration to 400 µSec.
Protocol VAR
- Timer is variable from 0 - 30 minutes, but Variable preset at 30 minutes, pulse rate is set to
400PPS, 65 µSec pulse duration. Pulse rate, timer and pulse duration are variable over their full range in this
mode.
Pulse Duration:
400, 200, 100, 65, 40, 20 µs available
Pulse Repetition Rate:
50, 100, 200, 400, 600, 800 PPS available
Timer:
The digital treatment timer indicates the set time in minutes and seconds prior to the start of treatment and treatment
time remaining during treatment. The maximum setting of the timer is 30 minutes. The timer increments in 1 or 10
minute units.
Timer accuracy:
+/- 1 second at all settings
Electrode:
Manually Tunable Hi Q Drum Inductive Applicator
Cooling:
Forced air by integrated fan
ENVIRONMENTAL CONDITIONS:
For Transport and Storage:
Temperature: -10 to 35oC
Relative Humidity: 5 to 95%
Atmospheric Pressure: 500 to 1060hPa
Each unit is supplied with a main power cable. Both the cable and the plug on the unit conform to international
safety standards.
All information on supply voltages, model, serial number, and month/year of manufacture is located on the rear
panel. The external fuse ratings and type information is located on the bottom of the unit.
Standards
The OmniSWD
®
has been designed to meet the requirements of IEC 60601-1:2005; “Safety of Medical Electrical
Equipment, Part l: General requirements.”
In addition the
OmniSWD
®
has been designed to meet the requirements of IEC 60601-2-3 “Specification for safety
for shortwave therapy equipment,” and the requirements of IEC 60601-1-2:2014 4
th
Edition “Collateral Standard:
Electromagnetic Compatibility: General requirements.”