User's Manual
For
Oil Pollution Detector
KLD-3
Please read instruction manual carefully before use
!
Страница 1: ...User s Manual For Oil Pollution Detector KLD 3 Please read instruction manual carefully before use ...
Страница 2: ...u have understood the content of the operation manual correctly and operate the product according to the described procedure Please contact your account manager in case of any questions when operating Operation should be done under the instruction of the technicians Do not disassembly the equipment by yourselves in order to avoid any man made damage ...
Страница 3: ...roduction 4 1 2 Technical Data 4 2 Operation 6 2 1 Hose Connection 6 2 2 Instrument Start 7 2 3 Measuring interface 7 2 4 Instrument Rinsing 8 2 5 Battery Recharging 8 3 Printer 9 4 Trouble Shooting 9 5 Notes 10 6 Packing list 11 7 Attached sheet 11 ...
Страница 4: ... total equipment downtime improve system security prolong the service life of equipment The equipment adopts color LCD screen visualization in digital and graphically display the oil pollution condition Can detect mineral oil synthetic oil or fat phosphate such as hydraulic oil lubricating oil transformer oil insulating oil turbine oil turbine oil gear oil engine oil aviation kerosene oil Can be w...
Страница 5: ...ticle size including 2µm 5µm 15µm 25 µm 50µm and 4µm c 6µm c 14µm c Pollution level range ISO 4406 0 24 NAS 1638 00 12 Working pressure difference low voltage 0 7 3 M Pa Working pressure long term 4 M Pa 24 hour short term 20 M Pa 30 min Working flow 50 300 ml min Oil viscosity 10 100 CST plus or minus 20 Sensitivity 1µm ISO4402 4µm c ISO11171 Repeatability precision 0 5 ISO Calibration period 1 y...
Страница 6: ...ruct how to use the KLD 3 to determine the level of oil contamination connect hoses software operation hose flushing etc Chart 3 0 Product panel 2 1 Hose Connection There are two ports labeled inlet and outlet The connecting methods are shown in Figure 3 0 below Ensure that connection of the nut to the hose is tighten with no oil leakage ...
Страница 7: ...cted in chapter 3 1 Open the cover plug in the power cable turn on the power switch to start the instrument Please see below for details 2 3 Measuring interface The default screen is as follows after the instrument has started Sampling period IOS Level ISO final grade The laser current Receiving voltage NAS Level NAS final grade ...
Страница 8: ...is the recommended fluid for flushing Insert the oil in hose in flushing fluid and the oil out hose into a waste container or bottle Select the rinsing button to start the pump Keep rinsing liquid flowing for a minute and then click the stop button Remove the hoses let the pump run for a few minutes to eliminate the remaining rinsing liquid in hoses Flushing now is completed Note For hose rinsing ...
Страница 9: ...rged 3 Printer This system printer is not a self locking button Each click of the print button can make the printer to print out the current test results final report generated in NAS1638 standard format The results are as shown in the figure below Open the cover install the thermal paper into place Close the cover leaving the end of the paper roll outside of the cover 4 Trouble Shooting Following...
Страница 10: ...ted The sample volume should be no less than 100ml this product is a precision laser instruments Strong vibration will Simple Fault Resolution for Fault Unable to start the instrument Check if the power cable is plugged in and power turned on Unable to print Check to see if the printer is short of paper No oil comes out from the oil out Check to see if the oil in hose is connected Check to see if ...
Страница 11: ...25 25 50 50 100 100 00 125 22 4 1 0 0 250 44 8 2 0 1 500 88 16 3 1 2 1000 178 32 6 1 No Item Specifications QTY 1 Online oil contamination detector ISO4406 NAS1638 1 2 Pressure hose M16 2 on both sides 1 5m 2 3 Imported PT type pressure joint M16x2 pressure connector 2 4 Electronic instruction CD 1 5 Paper instruction manual A4 1 6 Test report third party test report NAS standard 46 hydraulic oil ...
Страница 12: ... 4096 SHEET 2 Standard table of ISO 4406 contamination class Unit particle quantity 100ml Particle quantity per 100ml Class code particle quantity 100ml Class code Larger than Upper limit value Larger than Upper limit value 1300000 00 2500000 00 28 4000 8000 13 6400000 0 1300000 00 27 2000 4000 12 3200000 0 6400000 0 26 1000 2000 11 1600000 0 3200000 0 25 500 1000 10 8000000 1600000 0 24 250 500 9...
Страница 13: ...130000 250000 18 4 8 3 64000 130000 17 2 4 2 32000 64000 16 1 2 1 16000 32000 15 0 1 0 8000 16000 14 1 ...
Страница 14: ...ANG FENG KE WEI EQUIPMENT TECHNOLOGY CO LTD Tel 86 10 82866783 8185 Website www hangf com Email sales hangf com Fax 86 10 82600379 Post Code 100141 Address NO 6 Dacheng Rd West 4th Ring Fengtai District Beijing ...