7
Figure 3: Hand Control connector with latch being inserted into lift connector
Test the Hand Control
Perform a brief test to ensure proper connectivity:
1.
Turn on the lift as shown on page 8.
2.
Press any button on the Hand Control. The Hand Control screen lights up.
3.
Raise and lower the lift using the Up button
and Down button
.
4.
If the lift moves up and down,
the Hand Control is properly connected.
5.
If the lift does not move correctly,
double-check:
That the lift is on
That the Hand Control connector is properly connected
That the Hand Control does not display
Slack Detected
If there are still problems with the lift, contact your local Handicare authorized dealer
for service.
Mount the Hand Control
While the lift is not in-use, the hand control can be mounted on to the top of the lift using the
angled slider. Align the hand control with the slider on the cover, and slide the hand control down.
To remove, pull the hand control up and over until it is disengaged from the slider.
Turn On the Lift
Before using the AP-450 lift system, check the lift, track, and sling for any
unusual wear
or damage. Refer to the owner’s manual for each piece of
additional equipment to determine what should be checked. If anything looks
unusual, contact your local Handicare dealer or your local Handicare
business development manager before use. Failure to check the lift could
result in damage to the lift or serious injury to the operator or the individual
being lifted.
1.
Press any button on the Hand Control to turn the lift on. The indicator light on lift membrane
lights green and the display screen on Hand Control displays
[Ready]
.
2.
If the lift batteries require charging,
the indicator on the membrane lights orange, a beep
audible alarm sounds when the button is pressed, and the Hand Control displays the
message
[Low Battery]
.
3.
If the lift batteries are discharged to 25% of capacity,
the Hand Control displays the
message
[Low Battery]
and the lift continues to function until the battery status reaches
10-15%.
4.
If the lift batteries are discharged to 10-15% of capacity,
the Hand Control displays the
message
[Very low battery]
, the indicator on the lift membrane lights red, and a beep
audible alarm sounds when the button is pressed. The lift will not raise, but will lower and
the emergency lowering will still function.
Turn Off the Lift
The lift automatically shuts off after approximately 5 minutes of non-use.