DAD 141.1 Technical Manual, Rev. 2.5 – April 2015
Page 51 of 54
10.13.6. TL Trigger Level
[
Index 0x211E or 0x2400
]
This command defines a level for a rising edge trigger on the measurement signal. Setting range: 0 to 99999.
Master (PC / SPS) sends
Slave (DAD 141.1) responds
Meaning
TL
T+99999
Request: TL = 99999
TL 1000
OK
Setup: TL = 1000
In the example a new measurement cycle would automatically start, if the signal exceeds 1000 d (e.g. 100,0 g;
trigger commands SD and TL).
Default setting: TL = 99999 [= trigger level disabled]
Note:
All trigger possibilities are always available in parallel. If a software trigger (command TR) or a hardware
trigger (Digital input 0) will be used the trigger level should be set to its maximum value (TL = 99999). This
setting disables the trigger level.
Figure: Time plot of a typical checkweigher cycle
10.13.7. SA Send Triggered Average Value automatically
Master (PC / SPS) sends
Slave (DAD 141.1) responds
Meaning
SA
OK
Auto-Transmit: triggered average value
This command will start to auto-transmit the measurement value of the current trigger cycle.