DAD 141.1 Technical Manual, Rev. 2.5 – April 2015
Page 49 of 54
10.12. Save Calibration and Setup
– CS, WP, SS, AS, GI, PI
The calibration and setup parameters can be divided in 4 groups:
Calibration
: CM, DS, DP, CZ, CG, ZT, IZ and FD, etc. saved by command
CS
Setup
: FL, FM, NR, NT, BR, AD, DX and others, saved by command
WP
Setpoints:
S1, S2, S3, H1, H2, H3, A1, A2, A3 - saved by command
SS
Analog outputs:
AA, AH, AL, AM - if available - saved by command
AS
Note:
Calibration data can only be saved if the TAC code is known and preceding the CS command. See the
commands
CE
and
CS
in chapter 10.2.
The setup data and the setpoint data will be stored non-volatile in the EEPROM using the
WP
respective
SS
and
AS
command.
10.12.1. CS Save the Calibration Data
[
Index 0x2066
]
This command results in the calibration data being saved to the EEPROM and causes the TAC to be
incremented by 1.
Master (PC / SPS) sends
Slave (DAD 141.1) responds
Meaning
CE
E+00017 (example)
Request: TAC counter CE17
CE 17
OK
Calibration sequence active
CS
OK
Calibration values saved
The CS command saves all of the calibration group values, as set by AG, AZ, CZ, CG, CM, DS, DP, ZT etc.
The command returns ERR and has no updating action unless it is preceded by the CE_XXXXX.
10.12.2. WP Save the Setup Parameters
[
Index 0x2066
]
With this command the settings of the “Filter” (FL, FM), the “No-motion” (NR, NT) and the communication (AD,
BR, DX) will saved in the EEPROM.
Master (PC / SPS) sends
Slave (DAD 141.1) responds
Meaning
WP
OK
Setup data saved
WP
ERR
Error
10.12.3. SS Save Setpoint Parameters
[
Index 0x2066
]
With this command the setpoints (S0, S1), the setpoint hysteresis (H0, H1) and the setpoint allocation (A0, A1)
will be saved in the EEPROM.
Master (PC / SPS) sends
Slave (DAD 141.1) responds
Meaning
SS
OK
Setpoint parameters saved
SS
ERR
Error
10.12.4. AS Save Analogue Output Parameters
[
Index 0x2066
]
With this command the action (AA), the analog low (AL), the analog high (AH) and the output mode (AM) will be
saved in the EEPROM.
Master (PC / SPS) sends
Slave (DAD 141.1) responds
Meaning
AS
OK
Analog output parameters saved
AS
ERR
Error
10.12.5. GI Get an Image File from the EEPROM
Retrieves a HEX-INTEL formatted EEPROM image file from the EEPROM of the source DAD 141.1. The image
file contains all stored information except the calibration data. This image file can be downloaded to any
DAD 141.1 with the same firmware type and revision no. as the source DAD 141.1.
10.12.6. PI Download an Image File to the EEPROM
Downloads a HEX-INTEL formatted EEPROM image file to the target DAD 141.1 EEPROM. The image file
contains all stored information except the calibration data.
Attention:
The target DAD 141.1 must have same firmware type and revision no. as the source DAD 141.1.