DAD 141.1 Technical Manual, Rev. 2.5 – April 2015
Page 20 of 54
7.9. Menu 7
– Logic Outputs
Remark: Activate a new setup with 1x Power OFF & ON !
7
Logic output setup
(Menu 7.0 to 7.2)
7.0
Logic Output "0"
7.0.1
Setpoint "0"
7.0.1.1
Setup of the Setpoint value
(command S’n‘ –
n=0)
Permitted /- 999 999
7.0.1.2
Setup the Polarity (switch logic)
ON
or
OFF
(command P’n‘ –
n=0)
Use the UP/DOWN buttons for "on" / "oFF"
7.0.2
Hysterisis value "0" (± 'n')
(command H’n‘ –
n=0)
Permitted /- 9 999
7.0.3
Base for Setpoint "0"
(command A’n‘ –
n=0)
gros - G
ross
value
net - N
et
value
PEA - P
eak
value (Maximum)
AUEr - A
verage
value
HoLd - H
old
value
PP -
Peak to Peak
value
UALL -
Valley
value (Minimum)
Error -
Error
4
or
5
oFF - set output "1"
OFF
7.0.4
Test logic output "0"
(Use the UP/DOWN buttons)
Open/Close contacts using the keyboard
7.0.4.0
Output is
OFF
7.0.4.1
Output is
ON
7.1
Logic Output "1"
(commands S’n‘, P’n‘, H’n‘, A’n‘ –
n=1)
As per section 7.0 - but for logic output "1"
7.2
Logic Output "2"
(commands S’n‘, P’n‘, H’n‘, A’n‘ –
n=2)
As per section 7.0 - but for logic output "1"
7.3
Hold Time for all the Logic Outputs 0, 1 and 2
(command HT)
Permitted value range is from 0 to 65 535 ms
The signal has to exceed the setpoint limit continuously at least for this time
period before a switch event will be initiated.