Accumulated Scans
•
Indicates the total number of scans required (red indicators) for a
symbol profile; the number of scans already completed (green bars).
As scans are completed, red indicator bars turn green.
Right-click to set: Number of scans
•
ISO/IEC Symbol Grade
•
Running average of all scans, symbol grade in ISO/IEC format:
Alpha Grade/Numeric Grade/Aperture (mils)/Light Source
(nanometers)
Note:
Multiple scans are required for an ISO/IEC symbol grade. Multiple
scan profiles (and corresponding grades) are used to generate the symbol grade.
An individual scan profile is often useful for diagnostic purposes but cannot be
used as a complete judge of symbol quality. For more information on ISO/IEC
verification, see “Overview of ISO/IEC Verification.”
General Scan Results
•
Symbol Data
•
Characters decoded from the symbol (including special symbology
functions and control characters if enabled).
•
For messages longer than 32 characters, the scroll arrows allow you
to view the entire message.
Right-click the Symbol Data area to select whether to display
: Start/stop characters
: Check characters
: Code 128 modes (non-data characters)
: Expanded UPC-E0 message
•
Symbology Decoded
•
The symbology decoded is displayed here.
•
Pass/Fail indicator (pass indicates that an enabled symbology was
decoded, fail indicates that the symbology was not valid or not
enabled)
•
Visual Pass/Fail indicator (Pass/Fail background displays red or
green)
Right-click to select/deselect
:Symbologies
:EAN/UPC magnification factors and addendums.
•
Scan Direction
•
The allowable direction for scanning.
Right-click to select
14