45
HAMEG Instruments GmbH • Industriestr. 6 • D-63533 Mainhausen • Deutschland
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen, Dipl.-Ing. Andre Vander Stichelen • AG Offenbach am Main HRB 41200
1
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
[SENSe:]VOLTage[:DC]:NULL[:STATe]?
Returns the null function state of the DC V measurement function.
Return values:
<State>
1 | 0
1
ON - Null function of the DC V measurement function is activated
0
OFF - Null function of the DC V measurement function is disabled
[SENSe:]VOLTage[:DC]:NULL:VALue {
<
Value
>
|MIN|MAX}
Sets the null value of the DC V measurement function.
Parameters:
<Value>
-1000V to 1000V (adjustable in 1nV steps)
MIN: -1000V
MAX: 1000V
*RST: 0V
[SENSe:]VOLTage[:DC]:NULL:VALue? [{MIN|MAX}]
Returns the null value of the DC V measurement function.
Return values:
<Value>
-1000V to 1000V
MIN: -1.03
MAX: 1.03
[SENSe:]VOLTage[:DC]:RANGe:AUTO
<
Mode
>
Activates or disables the auto range mode of the DC V measurement function.
Parameters:
<Mode>
ON | OFF | 1 | 0
ON:
Activates the auto range mode of the DC V measurement function
OFF:
Disables the auto range mode of the DC V measurement function
1:
Activates the auto range mode of the DC V measurement function
0:
Disables the auto range mode of the DC V measurement function
[SENSe:]VOLTage[:DC]:RANGe:AUTO?
Returns the auto mode state of the DC V measurement function.
Return values:
<State>
1 | 0
1:
Auto range of the DC V measurement function is activated
0:
Auto range of the DC V measurement function is disabled