background image

Subject to change without notice

23

General

These Test Instructions are intended as an aid for checking
the most important characteristics of the HM303 at
regular intervals without the need for expensive test
equipment. Resulting corrections and readjustments inside
the instrument, indicated by the following tests, are
described in the Service Instructions or on the Adjusting
Plan. They should only be undertaken by qualified
personnel.
As with the First Time Operation instructions, care should
be taken that all knobs with arrows are set to their
calibrated positions. None of the pushbuttons should be
depressed. TRIG. selector switch to 

AC

.

 

It is recommended

to switch on the instrument for about 20 minutes prior to
the commencement of any check.

Cathode-Ray Tube: Brightness and Focus,
Linearity, Raster Distortions

Normally, the CRT of the HM303 has very good brightness.
Any reduction of this brightness can only be judged
visually. However, decreased brightness may be the
result of wrong setting or reduced high voltage. The latter
is easily recognized by the greatly increased sensitivity of
the vertical amplifier. Right setting means, that the 

HOLD

OFF

 control should be turned to the left stop; the 

X-MAG.

(x10)

 button should be released; a medium time coefficient

should be selected; line triggering (~ position) should be
used only with a suitable 

TIME/DIV.

 switch setting (e.g.

2ms/div.

). The control range for maximum and minimum

brightness (intensity) must be such that the beam just
disappears before reaching the left hand stop of the

INTENS.

 control (particularly when the 

XY

 button is

depressed), while with the control at the right hand stop
the focus and the line width are just acceptable.

With maximum intensity the timebase fly-back must
on no account be visible

. Visible trace fault without

input signal: bright dot on the left side  or  decreasing
brightness from left to right or shortening of the baseline.
(Cause: Incorrect Unblanking Pulse.) It should be noted
that with wide variations in brightness, refocusing is
always necessary. Moreover, with maximum brightness,
no “pumping” of the display must occur. If pumping does
occur, it is normally due to a fault in the regulation circuitry
for the high voltage supply. The presetting pots for the
high voltage circuit, minimum and maximum intensity, are
only accessible inside the instrument (see Adjusting Plan
and Service Instructions).

A certain out-of-focus condition in the edge zone of the
screen must be accepted. It is limited by standards of the
CRT manufacturer. The same is valid for tolerances of the
orthogonality, the undeflected spot position, the non-
linearity and the raster distortion in the marginal zone of

the screen in accordance with international standards
(see CRT data book). These limit values are strictly
supervised by HAMEG. The selection of a cathode-ray
tube without any tolerances is practically impossible.

Astigmatism Check

Check whether the horizontal and vertical sharpness of the
display are equal. This is best seen by displaying a square-
wave signal with the repetition rate of approximately 1MHz.
Focus the horizontal tops of the square-wave signal at
normal intensity, then check the sharpness of the vertical
edges. If it is possible to improve this vertical sharpness by
turning the 

FOCUS

 control, then an adjustment of the

astigmatism control is necessary. A potentiometer of 47k

is provided inside the instrument for the correction of
astigmatism (see Service Instructions). A certain loss of
marginal sharpness of the CRT is unavoidable; this is due to
the manufacturing process of the CRT.

Symmetry and Drift of the Vertical Amplifier

Both of these characteristics are substantially determined
by the input stages of the amplifiers.

The symmetry of both channels and the vertical final
amplifier can be checked by inverting Channel I and II
(depress the corresponding 

INVERT

 pushbutton). The

vertical position of the trace should not change by more
than 0.5div. However, a change of 1div. is just permissible.
Larger deviations indicate that changes have occurred in
the amplifier.

A further check of the vertical amplifier symmetry is
possible by checking the control range of the 

Y-POS.

controls. A sine-wave signal of 10-100kHz is applied to the
amplifier input. When the 

Y-POS.

 control is then turned

fully in both directions from stop to stop with a display
height of approximately 

8div.

, the upper and lower positions

of the trace that are visible should be approximately of the
same height. Differences of up to 1div. are permissible
(input coupling should be set to 

AC

).

Checking the drift is relatively simple. 

20 minutes after

switching on the instrument

, set the baseline exactly

on the horizontal center line of the graticule. The beam
position must not change by more than 

0.5div.

 during the

following hour.

Calibration of the Vertical Amplifier

Two square-wave voltages of 

0.2V

pp

 and 

2V

pp

 ±1% are

present at the output sockets of the calibrator (

CAL.

) If a

direct connection is made between the 

0.2V

 output and the

Test  Instructions

Содержание HM 303-4

Страница 1: ...ge 8 Amplitude Measurements 8 Time Measurements 9 Connection of Test Signal 10 First Time Operation 12 Trace Rotation TR 12 Probe compensation and use 12 Operating Modes of the Y Amplifier 14 X Y Oper...

Страница 2: ...ual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection to a cable Basica...

Страница 3: ...WG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied...

Страница 4: ...nal Features Component Tester 1kHz 1MHz Calibrator OSCILLOSCOPES Specifications Vertical Deflection Operating modes Channel I or II separate both Channels alternated or chopped Chopper frequency appro...

Страница 5: ...prox 500M DC 1MHz max 15kV DC peakAC HZ47 Viewing Hood for Oscilloscopes HM205 408 604 1 2 1005 and 1007 HZ48 Viewing Hood for Oscilloscopes 303 304 305 604 3 and 1004 during transpor tation of an osc...

Страница 6: ...is standard It has left the factory in a safe condition This instruction manual contains important information and warnings which have to be followed by the user to ensure safe operation and to retain...

Страница 7: ...tment designation isstatedforpossiblequeries thishelpstowards speeding up the processing of guarantee claims Maintenance Variousimportantpropertiesoftheoscilloscopeshouldbe carefully checked at certai...

Страница 8: ...However forsignalmagnitudesandvoltage designations in oscilloscope measurements the peak to peak voltage Vpp value is applied The latter corresponds to the real potential difference between the most...

Страница 9: ...he attenuator series resistor will break down causingdamagetotheinputoftheoscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal...

Страница 10: ...must be divided by 10 The fall time of a pulse can also be measured by using this method The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement tr...

Страница 11: ...obecompensation page M7 Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time In all cases where the oscilloscope bandwidth must be fully utilized e g...

Страница 12: ...ion required The HM303 accepts all signals from DC direct voltage up to a frequency of at least 30MHz 3dB For sinewave voltages the upper frequency limit will be 50MHz 6dB However in this higher frequ...

Страница 13: ...should then be 4 div 0 12div 3 During this adjustment the signal edges will remain invisible Adjustment at 1MHz Probes HZ51 52 and 54 can also be HF compensated They incorporate resonance de emphasing...

Страница 14: ...the 3 pushbuttons CH I II DUAL and ADD in the Y field of the front panel For Mono mode all 3 buttons mustbeintheirreleasedpositions onlychannelIcanthen beoperated ThebuttonCHI II TRIG I IImustbedepres...

Страница 15: ...in the X Y mode can exceed an angle of 3 above 120 kHz It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the...

Страница 16: ...o buttons Y CH I 20mV div AC TIME DIV 0 2ms div Triggering NORMAL with LEVEL setting internal or external triggering Figure 2 Amplitude modulated oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the...

Страница 17: ...ing mode it is possible to trigger at any amplitude point of a signal edge even with very complex signal shapes by adjusting the LEVEL control Its adjusting range is directly dependent on the display...

Страница 18: ...eringcanalsobeusefultodisplaysignalsbelowthe triggerthreshold lessthan0 5div Itisthereforeparticularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in...

Страница 19: ...iodes marked OVERSCAN which are located between the attenuators Should one LED illuminate without an input signal this means that the respective vertical positioning control has been improperly adjust...

Страница 20: ...orizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approxi mately evaluated The determination of actual values will come with...

Страница 21: ...n by 180 degrees round about the center point of the scope graticule In Circuit Tests Caution During in circuit tests make sure the circuit is dead No power from mains line or battery and no signal in...

Страница 22: ...ort circuit Resistor 510 Junction B C Junction B E Mains transformer prim Capacitor33 F Junction E C FET Z diode below 7V Z diode beyond 7V Diode paralleled by 680 2 Diodes antiparallel Silicon diode...

Страница 23: ...n in the edge zone of the screen must be accepted It is limited by standards of the CRT manufacturer The same is valid for tolerances of the orthogonality the undeflected spot position the non lineari...

Страница 24: ...anneldisplaywiththe CHI IIbutton is unnecessary it is contained indirectly in the tests above stated input of the vertical amplifier e g using a x1 probe the displayed signal in the50mV div position v...

Страница 25: ...splay should not shift horizontally during a change of the trigger coupling from AC to DC with a sine wave signal without DC offset If both vertical inputs are AC coupled to the same signal and both t...

Страница 26: ...LDOFF knob Component Tester After pressing the COMP TESTER button a horizontal straight line has to appear immediately when theCOMP TESTER socket is open The length of this trace should be approx 8div...

Страница 27: ...d from all voltage sources Normally the capacitors are discharged approx 6 seconds after switching off However with a defective instrument an interruption oftheloadisnotimpossible Therefore afterswitc...

Страница 28: ...g Checktoseethatallcircuitboardconnections are making good contact and are not shorting to an adjacent circuit Especially inspect the connections between the PCBs to front chassis parts to CRT PCB to...

Страница 29: ...ring TRIG EXT button depressed sync signal 0 3Vpp to 3Vpp to TRIG INP socket Line triggering TRIG selector switch to Select trigger coupling with TRIG selector switch Trigger frequency ranges AC 20Hz...

Страница 30: ...nector horizontal deflection in X Y mode Y POS II Controls vertical position knob of channel II display Inoperative in X Y mode POWER Turns scope on and off pushbutton LED LED indicates operating cond...

Страница 31: ...1 3 5 6 7 8 9 14 15 16 17 2 4 10 11 12 13 20 21 22 23 24 25 26 27 28 30 31 34 36 37 38 33 32 18 29 19 35...

Отзывы: