background image

Subject to change without notice

15

X-Y Operation

For 

X-Y operation

, the pushbutton in the X field marked

XY

 must be depressed. The X signal is then derived from

the 

INPUT CH II (X)

The calibration of the X signal

during X-Y operation is determined by the setting of
the Channel II input attenuator and variable control.

This means that the sensitivity ranges and input
impedances are identical for both the X and Y axes.
However, the 

Y-POS.II

 control is disconnected in this

mode. Its function is taken over by the 

X-POS.

 control. It

is important to note that the 

X-MAG. (x10)

 facility,

normally used for expanding the sweep, should not be
operated in the X-Y mode. It should also be noted that the
bandwidth of the X amplifier is 

3MHz (

3dB), and therefore

an increase in phase difference between both axes is
noticeable from 50kHz upwards.

The inversion of the X-input signal using the 

INVERT CH.II

button is not possible.

Lissajous figures can be displayed in the X-Y mode for
certain measuring tasks:

Comparing two signals of different frequency or bringing
one frequency up to the frequency of the other signal.
This also applies for whole number multiples or fractions
of the one signal frequency.

Phase comparison between two signals of the same
frequency.

Phase comparison with Lissajous figures

The following diagrams show two sine signals of the
same frequency and amplitude with different phase angles.

Calculation of the phase angle or the phase shift between
the X and Y input voltages (after measuring the distances

a

 and 

b

 on the screen) is quite simple with the following

formula, and a pocket calculator with trigonometric
functions. Apart from the reading accuracy, the signal
height has no influence on the result.

The following must be noted here:

Because of the periodic nature of the trigonometric
functions, the calculation should be limited to angles

90°. However here is the advantage of the method.

Do not use a too high test frequency. The phase shift
of the two oscilloscope amplifiers of the HM 303 in the
X-Y mode can exceed an angle of 3° above 120 kHz.

It cannot be seen as a matter of course from the screen
display if the test voltage leads or lags the reference
voltage. A CR network before the test voltage input of
the oscilloscope can help here. The 1 M

 input resis-

tance can equally serve as R here, so that only a
suitable capacitor C needs to be connected in series. If
the aperture width of the ellipse is increased (compared
with C short-circuited), then the test voltage leads the
reference voltage and vice versa. This applies only in
the region up to 90° phase shift. Therefore C should be
sufficiently large and produce only a relatively small just
observable phase shift.

Should both input voltages be missing or fail in the
X-Y mode, a very bright light dot is displayed on the
screen. This dot can burn into the phosphor at a too
high brightness setting (INTENS. knob) which causes
either a lasting loss of brightness, or in the extreme
case, complete destruction of the phosphor at this
point.

Phase difference measurement
in DUAL mode

A larger phase difference between two input signals of
the same frequency and shape can be measured very
simply on the screen in Dual mode (

DUAL

 button

depressed). The time base should be triggered by the
reference signal (phase position 0). The other signal can
then have a leading or lagging phase angle. Alternate
mode should be selected for frequencies 

1 kHz; the

Chop mode is more suitable for frequencies <1 kHz (less
flickering).
For greatest accuracy adjust not much more than one
period and approximately the same height of both signals
on the screen. The variable controls for amplitude and
time base and the 

LEVEL

 knob can also be used for this

adjustment  without influence on the result. Both base
lines are set onto the horizontal graticule center line with
the 

Y-POS

. knobs before the measurement. With

sinusoidal signals, observe the zero (crossover point)
transitions; the sine peaks are less accurate. If a sine
signal is noticeably distorted by even harmonics, or if a d.c.
voltage is present, 

AC

 coupling is recommended for 

both

channels. If it is a question of pulses of the same shape,
read off at steep edges.

a
b

2

(   )

cos

 ϕ

 ϕ

 ϕ

 ϕ

 ϕ

 

=

 

1

−−−−−

a
b

ϕϕϕϕϕ

 = arc sin

a
b

sin

 ϕ

 ϕ

 ϕ

 ϕ

 ϕ

 =

Содержание HM 303-4

Страница 1: ...ge 8 Amplitude Measurements 8 Time Measurements 9 Connection of Test Signal 10 First Time Operation 12 Trace Rotation TR 12 Probe compensation and use 12 Operating Modes of the Y Amplifier 14 X Y Oper...

Страница 2: ...ual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection to a cable Basica...

Страница 3: ...WG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied...

Страница 4: ...nal Features Component Tester 1kHz 1MHz Calibrator OSCILLOSCOPES Specifications Vertical Deflection Operating modes Channel I or II separate both Channels alternated or chopped Chopper frequency appro...

Страница 5: ...prox 500M DC 1MHz max 15kV DC peakAC HZ47 Viewing Hood for Oscilloscopes HM205 408 604 1 2 1005 and 1007 HZ48 Viewing Hood for Oscilloscopes 303 304 305 604 3 and 1004 during transpor tation of an osc...

Страница 6: ...is standard It has left the factory in a safe condition This instruction manual contains important information and warnings which have to be followed by the user to ensure safe operation and to retain...

Страница 7: ...tment designation isstatedforpossiblequeries thishelpstowards speeding up the processing of guarantee claims Maintenance Variousimportantpropertiesoftheoscilloscopeshouldbe carefully checked at certai...

Страница 8: ...However forsignalmagnitudesandvoltage designations in oscilloscope measurements the peak to peak voltage Vpp value is applied The latter corresponds to the real potential difference between the most...

Страница 9: ...he attenuator series resistor will break down causingdamagetotheinputoftheoscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal...

Страница 10: ...must be divided by 10 The fall time of a pulse can also be measured by using this method The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement tr...

Страница 11: ...obecompensation page M7 Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time In all cases where the oscilloscope bandwidth must be fully utilized e g...

Страница 12: ...ion required The HM303 accepts all signals from DC direct voltage up to a frequency of at least 30MHz 3dB For sinewave voltages the upper frequency limit will be 50MHz 6dB However in this higher frequ...

Страница 13: ...should then be 4 div 0 12div 3 During this adjustment the signal edges will remain invisible Adjustment at 1MHz Probes HZ51 52 and 54 can also be HF compensated They incorporate resonance de emphasing...

Страница 14: ...the 3 pushbuttons CH I II DUAL and ADD in the Y field of the front panel For Mono mode all 3 buttons mustbeintheirreleasedpositions onlychannelIcanthen beoperated ThebuttonCHI II TRIG I IImustbedepres...

Страница 15: ...in the X Y mode can exceed an angle of 3 above 120 kHz It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the...

Страница 16: ...o buttons Y CH I 20mV div AC TIME DIV 0 2ms div Triggering NORMAL with LEVEL setting internal or external triggering Figure 2 Amplitude modulated oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the...

Страница 17: ...ing mode it is possible to trigger at any amplitude point of a signal edge even with very complex signal shapes by adjusting the LEVEL control Its adjusting range is directly dependent on the display...

Страница 18: ...eringcanalsobeusefultodisplaysignalsbelowthe triggerthreshold lessthan0 5div Itisthereforeparticularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in...

Страница 19: ...iodes marked OVERSCAN which are located between the attenuators Should one LED illuminate without an input signal this means that the respective vertical positioning control has been improperly adjust...

Страница 20: ...orizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approxi mately evaluated The determination of actual values will come with...

Страница 21: ...n by 180 degrees round about the center point of the scope graticule In Circuit Tests Caution During in circuit tests make sure the circuit is dead No power from mains line or battery and no signal in...

Страница 22: ...ort circuit Resistor 510 Junction B C Junction B E Mains transformer prim Capacitor33 F Junction E C FET Z diode below 7V Z diode beyond 7V Diode paralleled by 680 2 Diodes antiparallel Silicon diode...

Страница 23: ...n in the edge zone of the screen must be accepted It is limited by standards of the CRT manufacturer The same is valid for tolerances of the orthogonality the undeflected spot position the non lineari...

Страница 24: ...anneldisplaywiththe CHI IIbutton is unnecessary it is contained indirectly in the tests above stated input of the vertical amplifier e g using a x1 probe the displayed signal in the50mV div position v...

Страница 25: ...splay should not shift horizontally during a change of the trigger coupling from AC to DC with a sine wave signal without DC offset If both vertical inputs are AC coupled to the same signal and both t...

Страница 26: ...LDOFF knob Component Tester After pressing the COMP TESTER button a horizontal straight line has to appear immediately when theCOMP TESTER socket is open The length of this trace should be approx 8div...

Страница 27: ...d from all voltage sources Normally the capacitors are discharged approx 6 seconds after switching off However with a defective instrument an interruption oftheloadisnotimpossible Therefore afterswitc...

Страница 28: ...g Checktoseethatallcircuitboardconnections are making good contact and are not shorting to an adjacent circuit Especially inspect the connections between the PCBs to front chassis parts to CRT PCB to...

Страница 29: ...ring TRIG EXT button depressed sync signal 0 3Vpp to 3Vpp to TRIG INP socket Line triggering TRIG selector switch to Select trigger coupling with TRIG selector switch Trigger frequency ranges AC 20Hz...

Страница 30: ...nector horizontal deflection in X Y mode Y POS II Controls vertical position knob of channel II display Inoperative in X Y mode POWER Turns scope on and off pushbutton LED LED indicates operating cond...

Страница 31: ...1 3 5 6 7 8 9 14 15 16 17 2 4 10 11 12 13 20 21 22 23 24 25 26 27 28 30 31 34 36 37 38 33 32 18 29 19 35...

Отзывы: