background image

Subject to change without notice

11

Sometimes the trace will disappear after an input signal
has been applied. The attenuator switch must then be
turned back to the left, until the vertical signal height is
only 3-8div. With a signal amplitude greater than 160V

pp

,

an attenuator probe must be inserted before the vertical
input. If, after applying the signal, the trace is nearly
blanked, the period of the signal is probably substantially
longer than the set value on the 

TIME/DIV.

 switch. It

should be turned to the left to an adequately larger time
coefficient.

The signal to be displayed can be connected directly to
the Y-input of the oscilloscope with a shielded test
cable such as HZ 32 or HZ 34, or reduced through a x10
or x100 attenuator probe. The use of test cables with
high impedance circuits is only recommended for
relatively low frequencies (up to approx. 50 kHz). For
higher frequencies, the signal source must be of low
impedance, i.e. matched to the characteristic resistance
of the cable (as a rule 50 Ohm). Especially when
transmitting square and pulse signals, a resistor equal
to the characteristic impedance of the cable must also
be connected across the cable directly at the Y-input of
the oscilloscope. When using a 50

 cable such as the

HZ 34, a 50

 through termination type HZ22 is available

from HAMEG. When transmitting square signals with
short rise times, transient phenomena on the edges
and top of the signal may become visible if the correct
termination is not used. A terminating resistance is
sometimes recommended with sine signals as well.
Certain amplifiers, generators or their attenuators
maintain the nominal output voltage independent of
frequency only if their connection cable is terminated
with the prescribed resistance. Here it must be noted
that the terminating resistor HZ22 will only dissipate a
maximum of 2 Watts. This power is reached with 10
Vrms or  at 28.3 V

pp

 with sine signal.

If a x10 or x100 attenuator probe is used, no termination
is necessary. In this case, the connecting cable is matched
directly to the high impedance input of the oscilloscope.
When using attenuators probes, even high internal
impedance sources are only slightly loaded (approx. 10
M

 II 16 pF or 100 M

 II 9 pF with HZ 53). Therefore, if

the voltage loss due to the attenuation of the probe can be
compensated by a higher amplitude setting, the probe
should always be used. The series impedance of the
probe provides a certain amount of protection for the input
of the vertical amplifier. Because of their separate
manufacture, all attenuator probes are only partially
compensated, therefore accurate compensation must be
performed on the oscilloscope (see “Probe compensation
page M7).

Standard attenuator probes on the oscilloscope normally
reduce its bandwidth and increase the rise time. In all
cases where the oscilloscope bandwidth must be fully

utilized (e.g. for pulses with steep edges) we strongly
advise using the 

modular probes HZ 51

 (x10) 

HZ 52

 (x10

HF) and 

HZ 54

  (x1 and x10. This can save the purchase

of an oscilloscope with larger bandwidth and has the
advantage that defective components can be ordered
from HAMEG and replaced by oneself. The probes
mentioned have a HF-calibration in addition to low
frequency calibration adjustment. Thus a group delay
correction to the upper limit frequency of the oscilloscope
is possible with the aid of an 1MHz calibrator, e.g. HZ60.

In fact the bandwidth and rise time of the oscilloscope are
not noticably changed with these probe types and the
waveform reproduction fidelity can even be improved
because the probe can be matched to the oscilloscopes
individual pulse response.

If a x10 or x100 attenuator probe is used, DC input
coupling must always be used at voltages above
400V

. With AC coupling of low frequency signals, the

attenuation is no longer independent of frequency,
pulses can show pulse tilts. Direct voltages are
suppressed but load the oscilloscope input coupling
capacitor concerned. Its voltage rating is max. 400 V
(DC + peak AC). 

DC

 input coupling is therefore of quite

special importance with a x100 attenuation probe which
usually has a voltage rating of max. 1200 V (DC + peak
AC). A 

capacitor

 of corresponding capacitance and

voltage rating may be connected in 

series with the

attenuator

 probe input for blocking DC voltage (e.g. for

hum voltage measurement).

With all attenuator probes, the 

maximum AC input voltage

must be 

derated

 with frequency usually above 20kHz.

Therefore the derating curve of the attenuator probe type
concerned must be taken into account.

The selection of the ground point on the test object is
important when displaying small signal voltages. It should
always be as close as possible to the measuring point. If
this is not done, serious signal distortion may result from
spurious currents through the ground leads or chassis
parts. The ground leads on attenuator probes are also
particularly critical. They should be as short and thick as
possible. When the attenuator probe is connected to a
BNC-socket, a BNC-adapter, which is often supplied as
probe accessory, should be used. In this way ground and
matching problems are eliminated.

Hum or interference appearing in the measuring circuit
(especially when a small deflection coefficient is used) is
possibly caused by multiple grounding because equalizing
currents can flow in the shielding of the test cables
(voltage drop between the protective conductor
connections, caused by external equipment connected to
the mains/line, e.g. signal generators with interference
protection capacitors).

Содержание HM 303-4

Страница 1: ...ge 8 Amplitude Measurements 8 Time Measurements 9 Connection of Test Signal 10 First Time Operation 12 Trace Rotation TR 12 Probe compensation and use 12 Operating Modes of the Y Amplifier 14 X Y Oper...

Страница 2: ...ual for a reduced cable length the maximum cable length of a dataline must be less than 3 meters long If an interface has several connectors only one connector must have a connection to a cable Basica...

Страница 3: ...WG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied...

Страница 4: ...nal Features Component Tester 1kHz 1MHz Calibrator OSCILLOSCOPES Specifications Vertical Deflection Operating modes Channel I or II separate both Channels alternated or chopped Chopper frequency appro...

Страница 5: ...prox 500M DC 1MHz max 15kV DC peakAC HZ47 Viewing Hood for Oscilloscopes HM205 408 604 1 2 1005 and 1007 HZ48 Viewing Hood for Oscilloscopes 303 304 305 604 3 and 1004 during transpor tation of an osc...

Страница 6: ...is standard It has left the factory in a safe condition This instruction manual contains important information and warnings which have to be followed by the user to ensure safe operation and to retain...

Страница 7: ...tment designation isstatedforpossiblequeries thishelpstowards speeding up the processing of guarantee claims Maintenance Variousimportantpropertiesoftheoscilloscopeshouldbe carefully checked at certai...

Страница 8: ...However forsignalmagnitudesandvoltage designations in oscilloscope measurements the peak to peak voltage Vpp value is applied The latter corresponds to the real potential difference between the most...

Страница 9: ...he attenuator series resistor will break down causingdamagetotheinputoftheoscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal...

Страница 10: ...must be divided by 10 The fall time of a pulse can also be measured by using this method The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement tr...

Страница 11: ...obecompensation page M7 Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time In all cases where the oscilloscope bandwidth must be fully utilized e g...

Страница 12: ...ion required The HM303 accepts all signals from DC direct voltage up to a frequency of at least 30MHz 3dB For sinewave voltages the upper frequency limit will be 50MHz 6dB However in this higher frequ...

Страница 13: ...should then be 4 div 0 12div 3 During this adjustment the signal edges will remain invisible Adjustment at 1MHz Probes HZ51 52 and 54 can also be HF compensated They incorporate resonance de emphasing...

Страница 14: ...the 3 pushbuttons CH I II DUAL and ADD in the Y field of the front panel For Mono mode all 3 buttons mustbeintheirreleasedpositions onlychannelIcanthen beoperated ThebuttonCHI II TRIG I IImustbedepres...

Страница 15: ...in the X Y mode can exceed an angle of 3 above 120 kHz It cannot be seen as a matter of course from the screen display if the test voltage leads or lags the reference voltage A CR network before the...

Страница 16: ...o buttons Y CH I 20mV div AC TIME DIV 0 2ms div Triggering NORMAL with LEVEL setting internal or external triggering Figure 2 Amplitude modulated oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the...

Страница 17: ...ing mode it is possible to trigger at any amplitude point of a signal edge even with very complex signal shapes by adjusting the LEVEL control Its adjusting range is directly dependent on the display...

Страница 18: ...eringcanalsobeusefultodisplaysignalsbelowthe triggerthreshold lessthan0 5div Itisthereforeparticularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in...

Страница 19: ...iodes marked OVERSCAN which are located between the attenuators Should one LED illuminate without an input signal this means that the respective vertical positioning control has been improperly adjust...

Страница 20: ...orizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approxi mately evaluated The determination of actual values will come with...

Страница 21: ...n by 180 degrees round about the center point of the scope graticule In Circuit Tests Caution During in circuit tests make sure the circuit is dead No power from mains line or battery and no signal in...

Страница 22: ...ort circuit Resistor 510 Junction B C Junction B E Mains transformer prim Capacitor33 F Junction E C FET Z diode below 7V Z diode beyond 7V Diode paralleled by 680 2 Diodes antiparallel Silicon diode...

Страница 23: ...n in the edge zone of the screen must be accepted It is limited by standards of the CRT manufacturer The same is valid for tolerances of the orthogonality the undeflected spot position the non lineari...

Страница 24: ...anneldisplaywiththe CHI IIbutton is unnecessary it is contained indirectly in the tests above stated input of the vertical amplifier e g using a x1 probe the displayed signal in the50mV div position v...

Страница 25: ...splay should not shift horizontally during a change of the trigger coupling from AC to DC with a sine wave signal without DC offset If both vertical inputs are AC coupled to the same signal and both t...

Страница 26: ...LDOFF knob Component Tester After pressing the COMP TESTER button a horizontal straight line has to appear immediately when theCOMP TESTER socket is open The length of this trace should be approx 8div...

Страница 27: ...d from all voltage sources Normally the capacitors are discharged approx 6 seconds after switching off However with a defective instrument an interruption oftheloadisnotimpossible Therefore afterswitc...

Страница 28: ...g Checktoseethatallcircuitboardconnections are making good contact and are not shorting to an adjacent circuit Especially inspect the connections between the PCBs to front chassis parts to CRT PCB to...

Страница 29: ...ring TRIG EXT button depressed sync signal 0 3Vpp to 3Vpp to TRIG INP socket Line triggering TRIG selector switch to Select trigger coupling with TRIG selector switch Trigger frequency ranges AC 20Hz...

Страница 30: ...nector horizontal deflection in X Y mode Y POS II Controls vertical position knob of channel II display Inoperative in X Y mode POWER Turns scope on and off pushbutton LED LED indicates operating cond...

Страница 31: ...1 3 5 6 7 8 9 14 15 16 17 2 4 10 11 12 13 20 21 22 23 24 25 26 27 28 30 31 34 36 37 38 33 32 18 29 19 35...

Отзывы: