34
Step Instructions
Visual Aid
21.
a. On the overlapped part of the poly strap, place
a
poly strap seal
.
b. Crimp poly strap seal at both ends of the seal
using the
poly strapping sealer
.
c. Repeat this step for each poly strap.
22.
Ship the pallet of TLS-EQ-77H systems to
Energetiq as air freight.
N/A
7.3. Fault Indicator Block Diagram
Figure 7:
Fault Indicator Block Diagram
Lamp Fault
Asserted
(pin 3 on remote)
Diagnosti
Safet
Bulb Interlock
Open
Lamp Head
Overtemp > 85C
Control Cable Not
Connected
Plasma
Extinguished
(after ignition)
Controller
Fault Asserted
(pin 4 on remote)
Laser Overtemp
(>60C)
Controller PCB
Overtemp (>82C)
Laser Current
Regulation Error
External
Interlock Open
Laser Temperature
Regulation Error
Signal Available on User Interface
Internal Logic That Generates Signal
Lamp On
De-Asserted
(pin 1 on remote)
Laser On
De-Asserted 150 sec
after Operate
Command
(pin 2 on remote)
Fail to Ignite
Failed Ignition
Содержание ENERGETIQ LDTLS TLS-EQ-77H