ORCA-Lightning Digital CMOS Camera C14120-20P Instruction manual_Ver.2.5
16
8-2 SET UP
First, set up the following before operating the camera.
(1) Setup of the computer and installation of an application software
Set up a computer to use beforehand, such as installing interface or application software.
Perform these procedures according to the each instruction manual.
(2) Preparing the mount adapter for attaching the lens
When using the F-mount, the lens can be attached to the camera in state of the product
shipment. The F-mount adapter is already attached at the product shipment.
- When using the F-mount lens, it is just attached.
- When using the C-mount, change to the dedicated adapter for the C-mount. (Refer to 8-4-1)
- When using the TFL-II-mount, detach the adapter attached to the camera. (Refer to 8-4-2)
When the mount part is ready, attach the same type of a lens or an optical system.
•
When using the C-mount, do not to screw the lens more than 7 mm from
the surface of the C-mount adapter on the camera. Doing so can scratch
the protective glass. (Some wide-angle lenses in particular can have a
thread of 7 mm or more.)
(3) Connecting to peripherals and outlets
Perform these connections according to 7. “CABLE CONNECTIONS”.
About CoaXPress interface connectors;
Each connector on the CoaXPress interface board and the camera has a number. When they are
connected, connect the same number of connectors to each other with CoaXPress interface
cables.
•
When cables are connected, turn off the power supply of the camera and
peripheral devices.
(4) When the camera is fixed by options
When the options such as Adjuster pole or Base plate are used and the camera is fixed, attach
according to procedure of each installation manual.