10-4 MODE SELECTOR
The mode selector allows choosing FOCUS mode or OPERATE mode.
(1) FOCUS mode
This mode is used during adjustment of the optics to check whether there is light
incident on the streak camera. The input light state is easy to check because no
sweep is performed. The incident light in FOCUS mode is not time-resolved, so the
photoelectrons from the photocathode tend to concentrate on one part of the
phosphor screen. Consequently, an abnormally large current flows to the MCP and
phosphor screen, which may damage the streak tube. During FOCUS mode, reduce
the incident light to a sufficient low level and then start adjusting the optical axis.
(2) OPERATE mode
This is for sweep mode and enables vertical sweep. Entering the trigger and
high-frequency signal sets the timing to begin a sweep.
10-5 GATE MODE
10-5-1 GATE FUNCTION
This system contains three gate modes. Gate functions operate when a voltage is applied to
the photocathode or MCP. Gate utilizing the MCP are called the MCP gate, and gate utilizing
the photocathode is called the photocathode gate. Select the gate mode according to the
measurement purpose and the type of light to measure. Gate modes that can be set differ
depending on the plug-in unit combinations. See chapter 11.Plug-in combinations and
operating conditions, to find out which modes are actually usable.
Содержание C10910
Страница 2: ......
Страница 89: ...C10910 INSTRUCTION MANUAL Ver 1 19E 87 Spacer ring for output optics ...
Страница 90: ...C10910 INSTRUCTION MANUAL Ver 1 19E 88 16 2 C10910 POWER SUPPLY UNIT Weight approx 5 9 kg Unit mm ...
Страница 92: ...C10910 INSTRUCTION MANUAL Ver 1 19E 90 16 4 M10911 02 SYNCHROSCAN UNIT Weight approx 3 85 kg Unit mm ...
Страница 94: ...C10910 INSTRUCTION MANUAL Ver 1 19E 92 16 7 M10913 11 SLOW SINGLE SWEEP UNIT Weight approx 3 4 kg Unit mm ...
Страница 95: ...C10910 INSTRUCTION MANUAL Ver 1 19E 93 16 8 M10914 01 HORIZONTAL BLANKING UNIT Weight approx 3 0 kg Unit mm ...
Страница 96: ...C10910 INSTRUCTION MANUAL Ver 1 19E 94 16 9 M10914 02 SHIFT BLANKING UNIT Weight approx 2 2 kg Unit mm operate ...
Страница 97: ...C10910 INSTRUCTION MANUAL Ver 1 19E 95 16 10 M10915 01 SYNCHRONOUS BLANKING UNIT Weight approx 2 85 kg Unit mm ...
Страница 99: ...C10910 INSTRUCTION MANUAL Ver 1 19E 97 16 12 A1976 01 INPUT OPTICS Weight approx 1 3 kg Unit mm ...
Страница 100: ...C10910 INSTRUCTION MANUAL Ver 1 19E 98 16 13 A1974 INPUT OPTICS Weight approx 2 1 Kg Unit mm depth 3 ...
Страница 102: ...C10910 INSTRUCTION MANUAL Ver 1 19E 100 16 16 A11771 01 MOUNT TABLE FOR ORCA R2 Weight approx 690 g Unit mm ...
Страница 103: ...C10910 INSTRUCTION MANUAL Ver 1 19E 101 16 17 A11771 03 MOUNT TABLE FOR ORCA FLASH4 0 Weight approx 1 0 kg Unit mm ...