halstrup-walcher P29 Скачать руководство пользователя страница 6

 

P29 Instruction Manual 

 

 

Safety precautions 

2.1 Safety concept 

When using in systems with natural gas, the entire pressure system must be flooded with the 
gas before switching on the electrical power supply. This creates a non-explosive atmosphere 
in the gas-carrying system of the unit. 

During operation, both pressure ports must be connected to the pressurized system. 

 
The P29 has been subjected to an ignition hazard analysis according to DIN EN 1127-1:2019. 
Under  normal  operating  conditions,  there  is  no  explosive  atmosphere  in  the  transducer. 
Consequently, it has no ATEX label, as it does not fall within the sphere of application covered 
by Directive RL 2014/34 /EU. 

2.2 Appropriate use 

The instrument may only be installed outside potentially explosive atmosphere.

 

 

The P29 is used to measure differential pressure, volumetric flow, mass flow and flow velocity 
using  the  orifice  plate.  It  may  be  used  in  systems  with  natural  gas  provided  that  the 
commissioning rules described in Section 3 are observed. 
 

The operator  of the device is responsible for supplying a pure (oxygen-free) fuel gas 
atmosphere  (natural  gas).  The  manufacturer  is  not  liable  for  damage  resulting  from 
improper or non-intended use. 

 
Always  observe  the  operating  requirements 

–  particularly  the  permissible  supply  voltage  – 

indicated on the rating plate and in the “Technical data” section of this manual. 
 
The  instrument  may  only  be  handled  as  indicated  in  this  manual.  Modifications  to  the 
instrument are prohibited. The manufacturer is not liable for damages caused by improper use 
or failure to follow these instructions. Violations of this type render all warranty claims null and 
void. 

2.3 Changing to the device 

The device may only be handled in accordance with these operating instructions. Modifications 
to  the  device  are  not  permitted.  The  manufacturer  is  not  liable  for  damage  resulting  from 
improper or non-intended use. In this case, the warranty claims also expire. 
 

If the device is used improperly or not  for its intended purpose, there is a risk of the 
device losing its explosion protection effect. 

 

2.4 Shipping, assembly, electrical connections and startup 

Assembly and the electrical connections should only be handled by trained professionals. They 
should be given proper training and be authorised by the operator of the facility. This applies 
particularly to systems that work with natural gas. Special start-up rules (Section 3) apply in 
this area. 
 
The following ignition hazards must be taken into account during installation: 
 

 

“Hot gases”

 The gas temperature of the natural gas must be within the permissible 

ambient temperatures of the unit. If necessary, ensure sufficient cooling or heating of 
the gas or mount the unit in a location where the natural gas has assumed the ambient 
temperature of the unit. 

Содержание P29

Страница 1: ...Pressure Transducer Type P29 for Natural Gas halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E Mail info halstrup walcher com Internet ww...

Страница 2: ...ion New 2 5 Services 3 1 flush revised New 3 3 leak test 3 3 shipping now 3 4 chapter 10 problem added All illustrations labelled New 10 1 Check fuse New 10 2 Replace fuse 9 Specifications overload ca...

Страница 3: ...Instrument description 11 4 1 Features 11 4 2 User interfaces 11 4 3 Internal ports and keys 12 4 4 Front view 13 4 5 Readout of volume mass flow and flow velocity 14 Zero point calibration cycle 15 5...

Страница 4: ...20 Zero point calibration 20 CS creep suppression 20 Lightning 20 Colour change 20 Brightness 20 Contrast 21 Read factory settings 21 Code 21 List of commands 22 Technical data 24 Trouble shooting Se...

Страница 5: ...efully It is especially important to read and follow the notes on starting up in systems that use flammable gases Contact the manufacturer if you do not understand any part of this instruction manual...

Страница 6: ...dicated on the rating plate and in the Technical data section of this manual The instrument may only be handled as indicated in this manual Modifications to the instrument are prohibited The manufactu...

Страница 7: ...ent errors may occur if the instrument is not kept protected from sunlight Specific safety precautions are given in individual sections of this manual 2 5 Troubleshooting maintenance service repairs d...

Страница 8: ...lead to bodily injury up to and including death if the corresponding instructions are not followed CAUTION This warns you of a potential hazard that could lead to significant property damage if corres...

Страница 9: ...fore connecting the instrument to the gas lines both pressure lines should be flooded with gas otherwise the instrument itself must be flooded with gas for a longer period of time Appropriate equipmen...

Страница 10: ...le pressure test For the leak test shut off the two pressure lines to the P29 and note the initial pressure display or output signal Observe the differential pressure for 1 minute The differential pre...

Страница 11: ...terface Relay Keyboard Analogue output signal figure 4 1 functional description 4 2 User interfaces The following four keys are available Menu UP Right Enter Key Meaning Measurement Mode Meaning Menu...

Страница 12: ...mass or volume flow Press the key to display the current value for the operating hours meter 4 3 Internal ports and keys figure 4 2 Internal connectors and push buttons Output signal Power supply figu...

Страница 13: ...5V 5V 2 0V 12 0V 5 0V 7 0V 4 N 24V starts external zero point calibration Power supply No Designation Description 11 GND Ground 12 24VDC Power supply 24VDC 13 RS232 interface for PC RS232 interface f...

Страница 14: ...s square rooted for example is the differential pressure value generated by a dynamic pressure probe at volume flow Analog current output signal 0 20 mA or 4 20 mA Analog voltage output signal 0 10 V...

Страница 15: ...e adjusted in the Settings menu Zero point calibration can also be started with the Enter key if the device is in Measurement mode The P29 does not respond to keys being pressed during calibration 5 1...

Страница 16: ...ing Warning Settings 7 1 Display The display unit can be selected with this menu item Press the and keys to select the unit you want and press Enter to confirm The type of display pressure flow rate e...

Страница 17: ...le Value Unit 7 2 2 1 Value This is used to set the volumetric flow value displayed at the maximum scaled pressure 0 is always used as the lower scale value on the volumetric flow display root extract...

Страница 18: ...f min km h 7 3 Warning INFORMATION For safety reasons relay outputs are not possible with the P29 However exceeding or falling below the switching thresholds can be signalled visually and acoustically...

Страница 19: ...eded With this parameter you can set the duration of this warning signal for each warning The maximum value here is 1000 ms continuous tone If both warnings are active the longer warning signal of the...

Страница 20: ...rameter to 0 deactivates automatic zero point calibration CS creep suppression This parameter specifies the value for creep suppression as a percentage If this value falls short of the measured pressu...

Страница 21: ...ble Read factory settings The factory settings are established in the instrument before delivery With this menu item they can be reactivated if required Code You can use this menu item to specify an a...

Страница 22: ...alSG Flow rate m s at max scale value ScalO Min 0 floating PRelai1 Switching pressure relay 1 Pa 120 120 of the measurement range floating RRelai1 Switching direction for Relay 1 1 decreasing 0 off 1...

Страница 23: ...01 2 0 1 3 0 2 4 0 5 5 1 unsigned int UnitD Unit display 0 Pa 1 hPa 2 kPa 3 mbar 4 mmH2O 5 mmHg 6 Psi 7 inH2O 8 inHg 9 m s 10 m h 11 kg s 12 kg min 13 kg h 14 m s 15 mph 16 f s 17 f min 18 km h unsig...

Страница 24: ...ssure 100 kPa for all measurement ranges Ambient conditions Medium Air non aggressive gases Operating temperature 10 C to 50 C Storage temperature 10 C to 70 C Electrical data Rated input approx 6 VA...

Страница 25: ...cknowledge 10 1 Checking the fuse If the device does not work and in the case of devices with a display nothing is shown on the display the fuse may be defective To check the fuse follow the steps bel...

Страница 26: ...following steps To change the fuse follow the steps below 1 Make sure that the device is no longer connected to the power supply 2 Pull the fuse upwards 3 Then reinsert the new fuse accordingly only u...

Страница 27: ...P29 Instruction Manual 27 Dimension drawing figure 11 1 dimensional drawing...

Страница 28: ...Pa mbar mmH2O mmHg psi inH2O inHg m s m h kg s kg min kg h m s mph f s f min km h Scaling Pressure Top xxx xx Bottom xxx xx Unit Pa hPa kPa mbar mmH2O mmHg psi inH2O inHg Vol Flow Value xxx xxxx Unit...

Страница 29: ...n Italian French Output 4 20 mA 0 20 mA 5 5V 0 10 V Filter xxxxx ms key tone on off Resolution max 0 01 0 1 0 2 0 5 1 Zero point calibration xx min CS x x Lighting Off White Blue Colour change On Off...

Страница 30: ...P29 Instruction Manual 30 Certificate of Conformity NEU_7100 004324_P29 docx 02 2023 Ts...

Страница 31: ...P29 Instruction Manual 31 Notes...

Страница 32: ...P29 Instruction Manual 32 Notes...

Отзывы: