008-00000-AR1-01-201011

1

AR-1 Argon

Calibration Light Source

Installation and Operation Instructions

Description

The AR-1 Argon Calibration Source is a wavelength calibration source for UV-VIS-Shortwave NIR 
spectrophotometric systems. The AR-1 produces Argon lines from 696-1704 nm, and is an ideal lamp 
to use when performing fast, accurate spectrometer wavelength calibrations in the NIR range.  

The AR-1 features an SMA 905 Connector for interfacing with our optical fibers. It operates with a 12 
VDC power supply (included with the unit) or 9V battery (not included).

The following sections detail the features of the AR-1 Argon Calibration Light Sources.

AR-1 Argon Calibration Light Source

Note

The AR-1 is NOT designed to operate as an excitation source in your experiments.
Spectral lines above 922 nm are not easily detectable by Ocean Optics spectrometers.