PRESSURE CONTROLLER operating manual
17
Navigation &
Customization
Main Menu
The Main Menu system is accessed by
pressing the
MENU
button from the Main
Display (
•
CONTROL
),
ABOUT
(
),
TARES
(below),
BASIC
CONFIG
(
), and
ADV
SETUP
(
) enter their menus.
•
MAIN (
) exits
to the main display.
Taring
Taring ensures accurate measurements
by giving the device a zero reference.
Auto-tare automatically tares when a
zero setpoint is set for a short time, but
auto-tare is not recommended for
pressure controllers
in most situations.
How to Manually Tare
1.
Ensure that there is no flow,
and that pressure is zero.
2.
MENU
→
TARE
→
TARE PRESS
, or
via serial command (
).
Gauge pressure tares must be
done with the controller open
to atmospheric pressure.
When to Tare
•
After significant changes in pressure
•
After dropping or bumping the pressure controller
•
After installing the controller in a different orientation
Main Menu
Tare Menu