Installation
January 2023
•
Rev. 0
• Standard Devices User Manual
8
Filters
When pressure drop is not a concern, use in-line sintered
filters to prevent large particulates from entering the flow
body of the controller. Suggested maximum particulate sizes
are as follows
Mass Flow:
• 5 microns
for devices with flow ranges ≤1 SCCM.
• 20 microns
for devices with flow ranges
between 1 SCCM and 1 SLPM.
• 50 microns
for devices with flow ranges ≥1 SLPM.
Liquid:
• 20 microns
for devices with flow ranges ≤100 CCM.
• 40 microns
for devices with flow ranges >100 CCM.
Pressure:
•
Maximum
40 microns
for all gauges.
• 20 microns
for PC, PCD, PCP, PCPD,
PCH, PCHD, PC3, PCD3
40 microns
for PCR, PCRD, PCR3, PCRD3
Power and Signal Connections
Power can be supplied to your device through either the
power jack or the multi-pin connector on top of your device.
For more pinout configurations, see
✓
Note:
Power requirements vary based on
analog configuration and valve type. Please
reference the associated specification sheet at
for power requirements.
⚠
Warning:
Do not connect power to pins 1
through 6, as permanent damage can occur. It
is common to mistake pin 2 (labeled 5.12 Vdc
Output) as the standard 0–5 Vdc analog output
signal. Pin 2 is normally a constant 5.12 Vdc.
Controller 8-Pin Mini-DIN Pinout
Meter and Gauge 8-Pin Mini-DIN Pinout
Pin
Function
Pin Function
1
Not Connected
Optional: 4–20 mA primary output signal
1
Not connected
Optional: 4–20 mA primary output signal
2
Static 5.12 Vdc by default.
Optional: secondary analog
output (4–20 mA, 0–5 Vdc, 1–5 Vdc, 0–10 Vdc) or alarm
2
Static 5.12 Vdc
Optional: secondary analog output
(4–20 mA, 0–5 Vdc, 1–5 Vdc, 0–10 Vdc) or alarm
3
Serial RS-232 RX / RS-485(
−
) Input Signal (receive)
3
Serial RS-232RX / RS-485(–) input signal (receive)
4
Analog Setpoint Input
4
Remote tare (ground to tare)
5
Serial RS-232 TX / RS-485(+) Output Signal (send)
5
Serial RS-232TX / RS-485(+) output signal (send)
6
0–5 Vdc
Optional: 1–5 Vdc or 0–10 Vdc output signal
6
0–5 Vdc
Optional: 1–5 Vdc or 0–10 Vdc output signal
7
Power In (as described above)
7
Power in (as described above)
8
Ground (common for power, digital
communications, analog signals, and alarms)
8
Ground (common for power, digital communications,
analog signals, and alarms)
The above pinouts are applicable to all devices with the mini-DIN connector. The availability of different output signals
depends on the options ordered. Optional configurations are noted on the unit’s calibration sheet.
Male Connector
Female Connector
1
4
5
6
8
7
3
2
1
4
5
6
8
7
3
2