Figure 4: Measure Screen of the HPC301
•
Push the
RUN/STOP
key
to begin a measurement. During the sampling process,
you may cancel the measurement by hitting the
RUN/STOP
key again.
•
The Start Delay will count down to zero, at which point the unit will begin sampling.
The unit will collect the number of samples requested at the appropriate Sample
Duration and Sample Interval.
•
If either the Sample Duration or Sample Interval is to be set below 5 seconds, the
other must be greater or equal to 5 seconds. For example, if you want a Sample
Duration of 4 seconds or less, the Sample Interval must be at least 5 seconds.
•
The instrument will automatically shut off after 5 minutes of inactivity unless
measurements are in progress.
Figure 5
outlines the HPC301 sampling process. The unit begins by counting trough
the start delay. This feature is useful if the unit will be programmed to take data up to 59
minutes and 59 seconds from the current time. If the start delay is set to 00:00,
sampling will begin immediately when the
RUN/STOP
key is pressed. The unit collects