Chapter 4, Getting Started
49
Haemonetics
®
TEG
®
6s User Manual
P/N 115191-IE, Manual revision: AB
Calibrate the
touchscreen
Haemonetics recommends that the touchscreen be calibrated upon first use or
installation. This ensures that the touch screen responds in the correct place.
It should not be necessary to calibrate the touchscreen more than once in the
lifetime of the analyzer.
To calibrate the touchscreen:
1. From the
Settings
screen, touch
calibrate
.
2. On the calibration screen, carefully touch and briefly hold a stylus on the
center of the target marked on the screen. Repeat this process as the
target moves around the screen.
Figure 4-17, Calibration screen
3. When no further targets are displayed, tap anywhere on the screen to
save the calibration and return to the
Settings
screen.
View
information
about the
analyzer
To view information about the analyzer:
1. From the
Settings
screen, touch
about
.
2. On the
About
screen, view information such as the version of the
software currently installed on the analyzer, serial number, IP address,
and the date and time that QC was last run.