Chapter 4, Getting Started
P/N 115191-IE, Manual revision: AB
Haemonetics
®
TEG
®
6s User Manual
Setting Up and Logging into the Analyzer
Set up the
analyzer
In most cases, a Haemonetics representative is responsible for the initial
unpacking and setup of the TEG analyzer. The following setup instructions are
provided in the event that a repaired analyzer is shipped back to you or that you
need to relocate the analyzer.
The analyzer ships with a US power cord that is suitable for use in North
America and Japan. Refer to the following table for a list of available
Haemonetics power cords.
Unpack the analyzer
1. From the shipping box, remove the small box containing the power
supply and extra ferrite clamp.
Note: The ferrite clamp is for installations where an Ethernet cable will be
permanently connected to the analyzer. For more information about
installing the clamp on the Ethernet cable and setting up a network
connection, contact TEG System Technical Support or refer to the TEG 6s
SMS Site Administrator Guide (116420-IE).
2. Remove the top foam supports and open the protective plastic bag.
3. Grasp the analyzer by the handle and lift it out of the shipping box.
Position the analyzer
Set up the TEG analyzer in location that is reasonably flat and level, and where
the analyzer will be protected from accidental damage. See
handling the analyzer” on page 24
for complete guidelines on how to correctly
place the analyzer.
Connect the analyzer to power
Screw the smaller end of the power cord onto the power jack at the back of the
analyzer, and then plug the other end into a grounded wall outlet.
Caution:
Always plug the power cord into the analyzer first, then into the
wall outlet. This prevents the possibility of electrical shorting or power supply
damage.
Region
Description
Part number
Europe
Power cord, AC, 4.6 meters (15 feet)
T1234-00
UK/Ireland
Power cord, 5 meters (16.4 feet)
106770-00
UK/Ireland
Power cord, 2.5 meters (8.2 feet)
85113-00
Switzerland
Power cord, 250V AC, 5 meters (16.4 feet)
85260-200