8-2
PCB Chassis Assembly Installation
P/N 108755-US, Manual revision: AA
I
NSTALLING
THE
PCB C
HASSIS
A
SSEMBLY
Caution:
Always wear an ESD strap when working on the Shaker.
Note: Models with serial number 0535 or greater are manufactured with an in-
terboard connector in place of the power supply to body board cable (P/N
10403300).
1.
Connect the extension harness (P/N 10405012-00) between the base and
PCB chassis assembly, as follows:
M = Motor connector, P3 on the Power Supply Board
SG = Strain Gauge Connector P2 on the Body Board
E = Reed Switch Connector, P3 on the Body Board
S = "Sonalert" Audio Alert Connector, P7 on the Body Board.
2.
Set the Shaker on its feet with the chassis assembly out of the base.
3.
Place AC ground wire directly from the input power receptacle onto the
chassis ground and secure with lock washer and nut.
4.
Connect Power Harness to P2 on the Power Supply Board.
5.
Connect head to body PCBA cable assembly to P1 on body board.
Warning: With the removal of the Body Board and Power Supply PCBA's from
the base, AC and DC voltages are exposed. Use CAUTION when handling these
assemblies.
Warning: Input Power fuse "F1" fuse holder, located on the power supply PCBA,
is exposed when the Body Board and Power Supply PCBA's are removed from
the base. Use CAUTI ON Awhen handling these assemblies.
Caution:
Ensure disconnected ground wires are controlled to prevent contact
with internal components.
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...