4-2
Ensuring Safety and Quality for a PCS2 Procedure
P/N 85266-30, Manual revision: A
H
ANDLING
THE
PCS2
EQUIPMENT
Safe and successful PCS2 operation will depend in part on the proper routine
handling of the PCS2 equipment. The operator should be aware of the problems
which could result if the device or disposable material is stored, installed or used
incorrectly.
Storing the PCS2
device and
material
The PCS2 device must not be operated or stored in an area where flammable
gases or vapors are present. The PCS2 disposable set material should be kept in
a dry, well-ventilated area and isolated from any chemical vapors. The operator
should handle the disposable set elements with clean, dry hands or gloves.
The ranges for storing the material should be within 8% to 80% rh and -20° C to
+50° C. The recommended conditions for the working environment are an
ambient temperature between 18° C to 27° C.
Inspecting the
material
Prior to installation, the operator should complete a visual inspection of the
disposable set elements and control for twisted or flattened sections.
After installing the disposable set, the operator should verify the correct place-
ment of the individual elements, prior to initiating a collection procedure. It is
important that the tubing remain free of any twists or occlusions which could
cause a flow obstruction.