4. Once the blue line has been emptied completely, remove the current
processing set from the device.
5. Install a new processing set, following the instructions beginning on
.
6. Once the new processing set is installed, touch
Resume Procedure
.
All procedure statistics from the procedure will be retained, and suction can
remain on throughout this process.
Changing the
Bowl Size
During a
Procedure
If the user selected the incorrect bowl size from the
Bowl Selection
screen,
navigate to the
Records
screen, view the procedure record, and edit the
processing set on the
Disposables
tab, following the instructions on