Automatic Operation: Cell Salvage
P/N 53063-50, Manual Revision: B
5-18
EMPTY BLUE
LINE Option
At the completion of the procedure, approximately 40 mL (13 mL for the 70 mL
bowl) of RBCs remain in the blue line that should be emptied to the reinfusion
bag. When the EMPTY mode is complete and prior to entering STANDBY a
message will flash for 10 seconds: “Is Case Completed? Y/N.”
If the NO key is pressed or no action is taken within 10 seconds, the machine will
enter STANDBY. If the YES key is pressed within 10 seconds, the following
message will flash: “Empty Blue Line? Y/N.”
If the YES key is pressed, the machine will empty the line at 100 mL/min and then
enter STANDBY. If the NO key is pressed, the machine will enter STANDBY
without emptying the line.
Warning: Do not empty the blue line during the procedure or the bowl
displacement air will be lost adversely affecting subsequent EMPTY modes.
Only empty the blue line at the completion of the procedure.
225
ML
RETURN
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
Press MODE twice to enter EMERGENCY option
AUTOMATIC
1250 ml
1000 ml
75 ml
0
Is Case Completed? Y/N
500 ml/min
225
ML
RETURN
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
Press MODE twice to enter EMERGENCY option
AUTOMATIC
1250 ml
1000 ml
75 ml
0
Empty Blue Line? Y/N
500 ml/min