Automatic Operation: Cell Salvage
P/N 53063-50, Manual Revision: B
5-16
Once the MODIFY key is pressed a screen similar to the following is displayed:
1. To turn on the NO WASH option use the SELECT key to highlight the “NO
WASH Option” and then press the up arrow to indicate ON.
2. Press MODIFY to return to the STANDBY screen.
Selection of the NO WASH option will cause a “Unit Not Washed” message to
appear on the screen during FILL.
Alternatively, you can move from FILL to EMPTY by just pressing the EMPTY key
once the machine trips into WASH.
Note: The operator must be aware that the processed blood will still contain
anticoagulant, increased free hemoglobin, and other cellular debris normally
removed by the washing process. The surgeon/anesthesiologist should be advised of
this deviation in processing and the reason for the variance documented.
STANDBY
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
to FILL automatically
Press START to FILL bowl or CONC to concentrate
Press MODIFY to change parameters
AUTOMATIC
0 ml
0 ml
0 ml
0
Need 800 ml in reservoir
CELL SAVER SETUP
Press ARROW keys to change values
Press SELECT to advance highlight
Min Wash Vol. - 225:
-125:
Reservoir Level:
Resume at Level:
Reservoir Level Sensor:
Pump Regulation:
Empty RBCs to:
AUTOMATIC
Press MODIFY when finished
NO WASH Option:
OFF
1000
750
800
400
ml
ml
ml
ml
ON
ON
Blue Line
Содержание Cell Saver 5+
Страница 2: ...Preface P N 53063 50 Manual Revision B ii...
Страница 134: ......