Automatic Operation: Cell Salvage
P/N 53063-50, Manual Revision: B
5-10
M
ODIFYING
THE
PRESET
PARAMETERS
The CS5+ is programmed with preset processing parameters that can be modified
by the operator while in AUTOMATIC operation. These parameters can all be
modified from the STANDBY state or some parameters can be changed at various
points in the processing cycle.
To modify the preset parameters the operator should:
1. Press the MODIFY key and a screen similar to the following will be
displayed:
2. Use the SELECT key to move from parameter to parameter. Use the UP/
DOWN arrows to adjust the values up or down or to toggle between
choices.
3. After changing the parameters, press MODIFY to return to STANDBY.
Note: All modifications to the preset parameters will be retained from procedure
to procedure until reprogrammed by the operator with the following exceptions:
1. A minimum wash volume below 1000 ml for a 225 ml bowl, 750 ml for
a 125 ml bowL will only be saved for the current procedure.
2. The NO wash option will only be saved for the current procedure.
3. The pump regulation off option (when available) will only be saved for
the current procedure.
4. The Protocol Parameter is not available if the 70 mL bowl adaptor has
been detected or the Cell Saver protocol has been entered.
CELL SAVER SETUP
Press ARROW keys to change values
Press SELECT to advance highlight
Min Wash Vol. - 225:
-125:
Reservoir Level:
Resume at Level:
Reservoir Level Sensor:
Pump Regulation:
Empty RBCs to:
AUTOMATIC
Press MODIFY when finished
NO WASH Option:
OFF
1000 ml
750 ml
800 ml
400 ml
ON
ON
Blue Line
Protocol:
Cell Saver