P/N 53063-50, Manual Revision: B
Automatic Operation: Cell Salvage
5-3
P
ROCESSING
IN
THE
AUTOMATIC
MODE
Before any blood can be processed, the disposables must be loaded correctly.
Follow the directions in the previous chapter.
Note: Anticoagulant, as prepared in Chapter 4, is not controlled by the CS5+.
This should be done manually in a 1:7 ratio (for heparin solution) of
anticoagulant rate to the rate of blood entering the reservoir.
1. After the system has been turned on, completed the self-test segment, shown
the current settings screen, and the disposables are properly loaded, the
following message is displayed:
The system is ready to begin processing when the fluid level in the collection
reservoir reaches the preset level.
Note: At any time during STANDBY the operator can initiate a FILL MODE by
pressing start or the backlit fill key when it is illuminated.
Note: The preset parameters can also be modified at this time, including adjusting
the preset level of the Reservoir level sensor, by pressing the MODIFY key.
Warning: Once the disposable is loaded and fluids connected, the operator
should not open the manifold over the clamps. This may result in a mixing of
fluids.
2. When the preset level is reached, the machine automatically enters the FILL
mode and the unprocessed blood will be pumped into the spinning
centrifuge bowl. On the first cycle (only) of a procedure, the saline line will
automatically be primed before the unprocessed blood is pumped into the
spinning centrifuge bowl.
STANDBY
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
to FILL automatically
Press START to FILL bowl or CONC to concentrate
Press MODIFY to change parameters
AUTOMATIC
0 ml
0 ml
0 ml
0
Need 800 ml in reservoir
Содержание Cell Saver 5+
Страница 2: ...Preface P N 53063 50 Manual Revision B ii...
Страница 134: ......
Страница 154: ......