Sequestering on the Cell Saver 5+ System
P/N 53063-50, Manual Revision: B
7-16
If the AIR bag was selected to manage the system air:
4. Spike the blood collection bag with either of the wash line bag spikes on the
Basic CS5+ Bowl Set. Close the clamp on the unused wash line spike.
5a. Press the START key to initiate filling the bowl from the blood bag. The
pump will start drawing at 60 ml/min.
b. As the bowl fills, plasma will be seen on the top followed by a white band
(platelets), and then red cells. The following screen will be shown:
STANDBY
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to fill bowl or CONC to concentrate
SEQUESTER
Press MODIFY to change parameters
PPP Volume:
PRP Volume:
0
0
0
0
0
ml
ml
ml
ml
0/
0/
Press YES to select Cell Saver protocol
CLOSE RED bag clamps
OPEN AIR bag clamp
CLOSE YELLOW bag clamp
CLOSE BLUE bag clamp
**AGITATE BLOOD BAG**
FILL BOWL
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press SELECT to advance highlight
SEQUESTER
90 ml
0 ml
0
Pump Rate:
60
Centrifuge:
4750
Press ARROW keys to change values
Press MODIFY to change parameters
ml/min
rpm
PPP Volume:
PRP Volume:
0 ml
0 ml
0/
0/