P/N 53063-50, Manual Revision: B
Sequestering on the Cell Saver 5+ System
7-7
M
ETHOD
1 — M
ANUAL
SEQUESTRATION
1. Collect whole blood from the patient via short intravenous or arterial cannula
into anticoagulated blood bag(s).
Warning: Only one unit of blood should be collected and processed at a time.
Unless directed by a physician, do not begin withdrawal of the second unit of
blood from the patient until the first unit of concentrated red autologous blood
cells has been returned.
2. Spike the blood collection bag with either of the wash line bag spikes on the
Basic CS5+ Bowl Set. Close the clamp on the unused wash line spike. On
the product bags, ensure that the RED bag clamps are closed and the white
bag clamps are open. On the harness, ensure the BLUE (PRP) bag clamp
and YELLOW (PPP) bag clamp are closed. Open the air bag clamp.
Note: If the PPP collection bag will be used for system air management, ensure
that the yellow bag clamp is open and the sir bag clamp is closed before starting
the procedure.
3a. Press START key to initiate filling the bowl from the blood bag. The pump
will start drawing at 60 ml/min.
b. As the bowl fills, plasma will be seen on the top followed by a white band
(platelets), and then red cells. The following screen will be shown:
4a. Observe the centrifuge bowl filling and allow air from the circuit and the
bowl to flow to the air bag (or PPP collection bag if used for air
management).
b. When plasma reaches the effluent line, open the yellow clamp to the PPP
bag and close the air bag clamp.
125
ML
FILL BOWL
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press SELECT to advance highlight
SEQUESTER
300 ml
0 ml
0
Pump Rate:
60
Centrifuge:
4750
Press ARROW keys to change values
Press MODIFY to change parameters
ml/min
rpm