BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
64
B.2.2 IEC 60601-1-2, Table 202
The BloodTrack Kiosk V5 is intended for use in the electromagnetic environment
specified below. The customer or operator of the BloodTrack Kiosk V5 should
ensure that it is used in such an environment.
IEC 60601-1-2, Table 202
Guidance
and manufacturer’s declaration―electromagnetic immunity
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete of
ceramic title. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
Transient/burst
IEC 61000-4-4
kV for power supply
lines
kV for input/
output lines
kV for power supply
lines
kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
kV line(s) to line(s)
kV line(s) to earth
kV line(s) to line(s)
kV line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5% U
T
(>95% dip in U
T
)
for 0,5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycle
70% U
T
(30% dip in U
T
)
for 25 cycle
< 5% U
T
(>95% dip in U
T
) for 5 s
< 5% U
T
(>95% dip in U
T
)
for 0,5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycle
70% U
T
(30% dip in U
T
)
for 25 cycle
< 5% U
T
(>95% dip in U
T
) for 5 s
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
BloodTrack Kiosk V5 requires
continued operation during power
mains interruptions, it is
recommended that the BloodTrack
Kiosk V5 be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
30 A/m
Power frequency magnetic field
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
❖
NOTE:
U
T
is the AC mains voltage prior to application of the level.