The pin details are:
Pin
Function
1
12V DC supply
2
Relay Common (RC)
3
Relay Normally Closed (NC)
4
Relay Normally Open (NO)
5
Input sensor
6
0V
Chassis
Shield via conductive P-clip
Table 1
Cabling
—pin details
Note:
The Input sensor is used
for the “door open” feature.
The BloodTrack Kiosk V5 emergency access button
—when engaged—will interrupt
the Relay Common (RC) connection. See
Section 8.3.8
The relay
—when enabled/energized—will connect the common (RC) to NO. When
the relay is off, the common (RC) will be connected to NC.
The 12V DC supply will be present all the time the BloodTrack Kiosk V5 is
connected to the mains power supply.
8.3.5 Digital Locks with Existing Interface
For storage locations that provide an existing lock interface (typically those fitted
with a digital lock access keypad), only the RC and NC or NO connections are used
as appropriate to the logic of the lock interface controller of the storage location. The
12V power supply connections are not used and neither is the use of any additional
diode (see below) required.
8.3.6 Powered Magnetic Lock Connections
A storage location fitted with a powered magnet lock
—see
Section 6.0
—will be connected as follows:
Number
Function
Connection
1
12V DC supply
To pin 2
2
Relay Common (RC)
From pin 1
3
Relay Normally Closed (NC)
Not connected
4
Relay Normally Open (NO)
To magnet
5
Sense Input
Door sensor (option)