BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
42
Figure 17
Electromagnetic lock
—connection to earth point
6.2.2 Mounting the Electromagnetic Lock
CAUTION
Extreme care should be taken to avoid drilling into glass edges.
The type of mounting hardware used will be decided by the installer, dependent
upon the construction of the storage location and any guidance given by the
manufacturer of the storage location. Suitable guidance from the manufacturer must
be obtained and followed.
Generally either self-tapping screws or .24 in. (6 mm) machine screws with drilled
and tapped holes will be used. Another option is using .16 in. (4 mm
) rivets. The ‘L’
bracket should use fixings that will allow positional adjustment to optimize closure of
the storage location.
Wherever possible, the electromagnetic locks should be located as close as
possible to the existing handle on the storage location
, or “pull point” to minimize
leverage.
Some modification of the brackets may be necessary. For example, shortening of
the bracket may be necessary, depending on the type of storage location and
available mounting position.
Notes:
•
Care should always be taken when drilling into cabinets and doors of the
storage location.
•
It is recommended that the bracket and fittings securing the magnetic lock be
inspected every six months.