
Preventive Maintenance and Installation
99
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
3. Enter Centrifuge Diagnostic Test mode.
4. Start the centrifuge by pressing
START
. Verify that the centrifuge speed
stabilizes to 8000 ± 100 rpm.
5. If calibration is required, refer to
.
6. As the centrifuge is spinning, observe the air/water interface in the bowl
and listen for any signs of centrifuge contact with the chassis. Verify that
the air/water interface never becomes discontinuous, indicating
excessive vibration. Listen for excessive or unusual noise in the
centrifuge. Run the centrifuge for 15 minutes to allow for misalignment to
show in the bowl header. If no noise or vibration is observed then the test
is complete and centrifuge is ok for use.
Figure 30, Example of a passing bowl vibration test.
Note: There can be a passing bowl vibration test that will not pass the test due
to high noise.
7. If excessive vibration or noise is observed due to centrifuge contact with
the chassis, center the centrifuge in the chassis by loosening the screws
that hold the base of the centrifuge underneath the machine and
retighten. After this, start the centrifuge speed, noise and vibration test
over from the beginning.
8. If the noise or excessive vibration cannot be resolved then replace
centrifuge.
9. Press
STOP
to stop the centrifuge and remove the bowl from the
centrifuge.
10.Exit the centrifuge test by pressing
MODIFY PROGRAM
.
Fluid Sensor
Test
Note: Power on the device in the Utilities mode of operation by pressing and
holding
SHAKER
while switching the power on. Hold
SHAKER
until the
UTILITIES menu appears on the display. On the UTILITIES menu, press
Air/fluid
interface
Содержание ACP 215
Страница 10: ......
Страница 38: ......
Страница 58: ......
Страница 104: ...104 Preventive Maintenance and Installation P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual...
Страница 126: ......
Страница 150: ......
Страница 182: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 182 System Schematics System Schematic...
Страница 183: ...Haemonetics ACP 215 Service Manual P N SM ACP215 01 EN AB System Schematics 183 Family Tree...
Страница 184: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 184 System Schematics Pneumatic Schematic...
Страница 186: ......