Disassembly
149
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Drain Tube Assembly
Drain Tube
1. Tilt the cabinet and remove the (4) Phillips head screws securing the spill
bag plate to the base of the cabinet.
2. Carefully cut the tubing from its mount.
3. Replace with new drain tube assembly.
Shaker
The shaker unit is not repairable in the field.
For technical assistance, or to have your shaker serviced in the USA, call
Haemonetics Customer Care Center at 1-800-537-2802.
For technical assistance, or to have your shaker serviced in other countries,
please contact your local Haemonetics representative.