9.2 Clean spills
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
1.
Obey all facility safety protocols for spill control.
2.
Discard the waste according to applicable regulations.
9.3 Clean the instrument
N O T I C E
Do not use solvents to clean the instrument.
The instrument is maintenance free. Regular cleaning is not necessary for normal operation. If the
exterior of the instrument becomes dirty, wipe the instrument surfaces with a clean, moist cloth.
9.4 Clean the vial
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
When the turbidity reading shows that there is contamination in the process vial or "VIAL CLARITY"
shows on the controller display, clean the vial.
1.
Push
menu
.
2.
Select SENSOR SETUP>TU5x00 sc>DIAG/TEST>MAINTENANCE>VIAL CLEANING.
3.
Complete the steps that show on the controller display. The instrument automatically saves the
cleaning process date after the last screen shows.
4.
If the optional automatic cleaning module is installed, push
menu
and select
SETUP>TU5x00 sc>START WIPE to start the automatic cleaning process.
5.
If the optional automatic cleaning module is not installed, clean the vial with the manual vial wiper.
N O T I C E
Carefully remove most of the water in the vial. Carefully put the vial wiper into the process vial so that no
water spills out.
Clean the process vial with the manual vial wiper as shown in the illustrated steps that follow.
English
23
Содержание TU5300 sc
Страница 12: ...12 English ...
Страница 15: ...English 15 ...
Страница 19: ...English 19 ...
Страница 27: ...English 27 ...
Страница 41: ...Français 41 ...
Страница 44: ...44 Français ...
Страница 49: ...Français 49 ...
Страница 57: ...Français 57 ...
Страница 71: ...Español 71 ...
Страница 74: ...74 Español ...
Страница 79: ...Español 79 ...
Страница 87: ...Español 87 ...
Страница 102: ...102 Português ...
Страница 105: ...Português 105 ...
Страница 110: ...110 Português ...
Страница 118: ...118 Português ...
Страница 132: ...132 中文 ...
Страница 134: ...图 4 检查干燥剂盒 1 安装到期日 mm yyyy 月份和 年份 2 指示灯 淡蓝色 未过期 白色 过期 3 运输安全保护 134 中文 ...
Страница 135: ...中文 135 ...
Страница 139: ...中文 139 ...
Страница 146: ...146 中文 ...
Страница 147: ...9 7 更换干燥剂盒 控制器屏幕将显示何时应更换干燥剂盒 请参阅干燥剂盒包装袋中随附的文档更换干燥剂盒 中文 147 ...
Страница 159: ...日本語 159 ...
Страница 161: ...図 4 乾燥剤カートリッジの検査 1 設置有効期限 mm yyyy 月と 年 2 インジケーター 水色 有効期 限内 白 有効期限切れ 3 輸送安全保護 日本語 161 ...
Страница 162: ...162 日本語 ...
Страница 167: ...日本語 167 ...
Страница 175: ...日本語 175 ...
Страница 188: ...188 ไทย ...
Страница 191: ...ไทย 191 ...
Страница 196: ...196 ไทย ...
Страница 204: ...204 ไทย ...
Страница 217: ...한글 217 ...
Страница 219: ...그림 4 건조제 카트리지 검사 1 설치 날짜 mm yyyy 월 및 연도 2 표시등 연청색 만료되지 않음 흰색 만료됨 3 운송 안전 보호 한글 219 ...
Страница 220: ...220 한글 ...
Страница 225: ...한글 225 ...
Страница 233: ...한글 233 ...
Страница 246: ...246 中文 ...
Страница 248: ...图 4 檢查乾燥劑濾芯 1 安裝有效期限 mm yyyy 月份 和年份 2 指標 淺藍色 未過期 白色 過期 3 運輸安全防護 248 中文 ...
Страница 249: ...中文 249 ...
Страница 253: ...中文 253 ...
Страница 258: ...9 5 清潔玻璃瓶分隔室 只有在分隔室受到汙染時需要清潔玻璃瓶分隔室 確認用來清潔玻璃瓶分隔室的工具表面柔軟 不會 損壞儀器 表 3 和图 7 顯示玻璃瓶分隔室清潔方式的選項 258 中文 ...
Страница 260: ...260 中文 ...
Страница 261: ...9 7 更換乾燥劑濾芯 控制器顯示幕會顯示乾燥劑濾芯的更換到期日 請參閱乾燥劑濾芯袋中隨附的文件 瞭解如何更換乾 燥劑濾芯 中文 261 ...
Страница 264: ......
Страница 265: ......