5.5 Replace the wiper
To make sure that the vial is cleaned fully, replace the wiper periodically.
1.
Push
menu
.
2.
Select SENSOR SETUP>[select analyzer]>DIAG/TEST>MAINTENANCE>WIPER REPLACE.
3.
Stop the sample flow.
4.
Remove the cleaning module.
5.
Remove the vial. Refer to steps 1 to 5 of
6.
Complete the steps that show on the controller display. Install the vial wiper (silicone or fiber) that
is applicable to the sample type. Refer to the illustrated steps that follow.
The date that the wiper was replaced is automatically saved after the last screen shows.
7.
Install the vial. Refer to steps 8 to 12 of
on page 14.
English
17
Содержание TU5 Series
Страница 2: ...Table of Contents English 3 Français 19 Español 36 Português 53 中文 70 日本語 86 한글 103 ไทย 119 2 ...
Страница 9: ...English 9 ...
Страница 10: ...10 English ...
Страница 11: ...English 11 ...
Страница 16: ...16 English ...
Страница 25: ...Français 25 ...
Страница 26: ...26 Français ...
Страница 27: ...Français 27 ...
Страница 32: ...32 Français ...
Страница 42: ...42 Español ...
Страница 43: ...Español 43 ...
Страница 44: ...44 Español ...
Страница 49: ...Español 49 ...
Страница 50: ...50 Español ...
Страница 59: ...Português 59 ...
Страница 60: ...60 Português ...
Страница 61: ...Português 61 ...
Страница 66: ...66 Português ...
Страница 76: ...76 中文 ...
Страница 77: ...中文 77 ...
Страница 78: ...78 中文 ...
Страница 83: ...中文 83 ...
Страница 92: ...92 日本語 ...
Страница 93: ...日本語 93 ...
Страница 94: ...94 日本語 ...
Страница 99: ...日本語 99 ...
Страница 102: ...アクセサリー 説明 数量 アイテム番号 マイクロファイバー布 バイアル洗浄用 1 LZY945 サービスブラケット 1 LZY873 チューブ TU5x00 sc の入口と出口 インチ 外径 4m LZY911 102 日本語 ...
Страница 109: ...한글 109 ...
Страница 110: ...110 한글 ...
Страница 111: ...한글 111 ...
Страница 116: ...116 한글 ...
Страница 125: ...ไทย 125 ...
Страница 126: ...126 ไทย ...
Страница 127: ...ไทย 127 ...
Страница 132: ...132 ไทย ...
Страница 133: ...ไทย 133 ...
Страница 136: ......
Страница 137: ......