11.
Two options are available during the procedure:
Option Description
Stop
Aborts the calibration and no results are calculated. If selected during the
Replicate
Sample
option, then all data in the series is lost.
Skip
Stops the current operation and goes directly to the next step in the procedure.
Results are then calculated from calibration data available before
Skip
is selected.
Results can be less accurate using this option.
12.
Push
Reject
or
Continue
.
Option
Description
Reject
Rejects the calibration. Select
Cancel
to go back to the result display or
Confirm
to reject the calibration. If this is the first calibration, select
Confirm
to reject the
calibration and use the default or previous calibration values. If this is a
Replicate
Standard
calibration, select
Confirm
to reject only the current calibration in the
series.
Continue
Select from one of these options:
•
Replicate Standard
: Do the calibration again using the same standard
•
Save & Exit
: Keep the calibration results and exit the calibration procedure
•
Reject & Exit
: Reject the calibration results and use the default or previous
calibration values, and exit the calibration procedure
7.1.3 Auto leveling calibration
This option is only available when at least one installed application contains an auto
leveling calibration method and the method is set to active. The calibration makes sure
that the sample volume in the measurement cell is the same before each titration.
An external pump must be installed for this procedure.
1.
From the home screen, select
Calibration
, then push
Autoleveling calibration
.
2.
If more than one application contains an auto leveling calibration method, push the up
and down arrow keys to highlight the application to use, then push
Select
.
3.
Calibration information shows on the display. If necessary, select an icon for more
information or to change some data.
Option
Description
Information
Shows more information about the calibration.
Operator
Changes the operator ID. Select from a list of applicable operators.
4.
Make sure that the external pump and tubes are correctly installed (refer to
on page 19).
5.
Make sure that the sensor, tubes and stir bar are correctly installed in an empty
beaker, as if preparing for a standard measurement.
6.
Accurately measure the necessary volume of sample and pour it into the empty
beaker.
7.
Adjust the position of the tube from the external pump (refer to
so that it touches the surface of the liquid. Then add more sample until the end of the
tube is immersed in 5 to 10 mm of the liquid.
8.
Push
Start
to start the calibration.
9.
The leveling procedure starts and sample is removed for the preset time. Make sure
that this time is large enough for all the necessary sample to be removed and for air
to be removed for some seconds at the end.
Standard operations
35
Содержание TitraLab AT1102
Страница 1: ...DOC022 52 93074 TitraLab AT1000 series workstations User Manual 03 2019 Edition 7...
Страница 2: ......
Страница 6: ...Specifications 4...
Страница 20: ...Installation 18...
Страница 25: ...Install accessories 23...
Страница 26: ...Install accessories 24...
Страница 34: ...Startup 32...
Страница 42: ...Standard operations 40...
Страница 50: ...Maintenance 48...
Страница 58: ...Troubleshooting 56...
Страница 64: ...Replacement parts and accessories 62...
Страница 74: ...Change application parameters 72...
Страница 79: ......