51750-23

Laboratory pH Meter Quick Reference Card

Changing Meter Setup Options

Setup options 1–5 have default settings (underlined in the table). To change the default setting, use
the table below. Settings for options 6–8 must be entered from the setup menu.

Setup

How To Get There

Description of Option

1

Turn the meter on and press
SETUP. Press ENTER.

Choose between using the BNC or 5-pin connector. ENTER
toggles between the options. Press EXIT when selection is
complete.

2

Turn the meter on and press
SETUP

.

Press the up arrow once.

Press ENTER.

Turn display lock On or Off. ENTER toggles between the
options. Press EXIT when selection is complete.

3

Turn the meter on and press
SETUP

.

Press the up arrow twice.

Press ENTER.

Choose between °C or °F for displayed temperatures.
ENTER toggles between the options. Press EXIT when
selection is complete.

4

Turn the meter on and press
SETUP

.

Press the up arrow three

times. Press ENTER.

Select measurement resolution from tenths, hundredths, or
thousandths. ENTER toggles between the options. Press
EXIT when selection is complete.

5

Turn the meter on and press
SETUP

.

Press the up arrow four

times. Press ENTER.

Select pH 6.86 or 7.00 as the pH the meter will automatically
recognize during calibration. ENTER toggles between the
options. Press EXIT when selection is complete.

6

Turn the meter on and press
SETUP

.

Press the up arrow five

times. Press ENTER

.

Select the time of day by pressing the desired number key.
The meter uses a 24-hour clock. All four places must have a
value. Use a 0 if the left-most digit does not have a value
(i.e., 03:08). Press ENTER to accept the value.

7

Turn the meter on and press
SETUP

.

Press the up arrow six

times. Press ENTER

.

Set the month and day by pressing the desired number
keys. The month uses the two places left of the slash; the
day of the month occupies the two places right of the slash.
Enter a 0 if the left-most digit does not have a value
(i.e., 02/06). Press ENTER to accept the value.

8

Turn the meter on and press
SETUP

.

Press the up arrow seven

times. Press ENTER

.

Set the year by using the number keys. Every place must
have a value. Press ENTER to accept the value.