b.
Use the data structure to group the data types. Go to
PLC data types
. Select
Add new data
type
and add a name for the data type ("SC_data_Type" in the example). Add the Tags and
variables based on the sensor values.
On the
Default tag table
, add the data structure and select the data type name. Add the
address of the data structure.
11.
Select the option
Monitor all
in the popup menu in the Default table screen. If the connection is
successful, the data from the SC4500 controller shows.
5.4.2 Configure the EtherNet/IP module
Configure the EtherNet/IP and the Telegram data in the controller to send sensor or device data to
the PLC. Refer to
EtherNet/IP Telegram composition
on page 40.
Configure the EtherNet/IP settings as follows:
1.
Push the main menu icon, then select
Outputs
>
EtherNet/IP
>
Telegram
.
English
31