Installation
3.
Screw a new sensor connector into the housing and connect the sensor connector to
a sensor connection on the main circuit board. Any available sensor connector can be
used, refer to
4.
Reassemble the housing.
3.13 Connecting AC Powered sc-Sensors
Note: AC power outlets can only be connected when a 100 V–240 V power supply is fitted in the
sc1000 controller system. To install the power outlets refer to
.
Most sc-sensors draw power directly from the sc-sensor connection. However,
certain sc-sensors may require supplemental 100–240 VAC power (e.g., to run pumps or
heating elements). These AC powered sc-sensors have two cables that connect to an
sc1000 probe module: a standard sc-sensor connector and a special connector for
drawing AC power from the probe module.
To connect AC powered sensors to a probe module follow these steps:
1.
Unscrew the cover on the AC power outlet.
2.
Connect the power connector from the analysis instrument to one of the
AC power outlets.
3.
Connect the sc-sensor connector to any available sc-sensor socket.