Reset calibration options
The calibration options can be reset to the factory default options.
1.
Push the
MENU
key and select Sensor Setup, [Select Sensor],
Calibrate, Reset Default Cal.
2.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
3.
Push
ENTER
. The Reset Cal? screen is shown.
4.
Push
ENTER
. All calibration options are set to the default values.
5.
If the option for operator ID is set to Yes in the Calibration Options
menu, enter an operator ID. Refer to
on page 14.
6.
On the New Sensor screen, select whether the sensor is new:
Option Description
Yes
The sensor was not calibrated previously with this controller. The
days of operation and previous calibration curves for the sensor are
reset.
No
The sensor was calibrated previously with this controller.
7.
Push the
BACK
key to return to the measure screen.
Modbus registers
A list of Modbus registers is available for network communication. Refer
to the manufacturer's website for more information.
Maintenance
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Clean the sensor
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
W A R N I N G
Personal injury hazard. Removal of a sensor from a pressurized vessel can be
dangerous. Installation and removal of these sensors should be done by
individuals trained in proper high pressure and temperature installation. Always
use industry approved hardware and safety procedures when dealing with high
pressure and/or temperature fluid transport systems.
Pre-requisite:
Prepare a mild soap solution with warm water and
dishwashing detergent, Borax hand soap or a similar soap.
Examine the sensor periodically for debris and deposits. Clean the
sensor when there is a buildup of deposits or when performance has
degraded.
1.
Use a clean, soft cloth to remove loose debris from the end of the
sensor. Rinse the sensor with clean, warm water.
2.
Soak the sensor for 2 to 3 minutes in the soap solution.
3.
Use a soft bristle brush to scrub the entire measuring end of the
sensor.
4.
If debris remains, soak the measuring end of the sensor in a dilute
acid solution such as < 5% HCl for a maximum of 5 minutes.
5.
Rinse the sensor with water and then return to the soap solution for
2 to 3 minutes.
6.
Rinse the sensor with clean water.
Always calibrate the sensor after maintenance procedures are done.
English
15
Содержание Polymetron
Страница 2: ...English 3 Français 20 Español 38 Português 58 中文 76 日本語 92 한글 109 ไทย 125 2 ...
Страница 8: ...1 2 3 4 8 English ...
Страница 9: ...5 6 7 8 English 9 ...
Страница 25: ...1 2 3 4 Français 25 ...
Страница 26: ...5 6 7 8 26 Français ...
Страница 42: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT 42 Español ...
Страница 44: ...1 2 3 4 44 Español ...
Страница 45: ...5 6 7 8 Español 45 ...
Страница 63: ...1 2 3 4 Português 63 ...
Страница 64: ...5 6 7 8 64 Português ...
Страница 79: ...图 3 安装示例 1 流通室 T 型安装 聚氯乙烯 英寸 NPT 2 流通室 T 型安装 不锈钢 1 4 英寸 NPT 中文 79 ...
Страница 81: ...1 2 3 4 中文 81 ...
Страница 82: ...5 6 7 8 82 中文 ...
Страница 95: ...図 3 取り付け例 1 フロースルー T 字取り付け PVC 19 mmNPT 2 フロースルー T 字取り付け ステンレス鋼 6 mmNPT 日本語 95 ...
Страница 97: ...1 2 3 4 日本語 97 ...
Страница 98: ...5 6 7 8 98 日本語 ...
Страница 112: ...그림 3 장착 예 1 흐름 통과 T 장착부 PVC 3 4 인치 NPT 2 흐름 통과 T 장착부 스테인리스 스틸 1 4 인치 NPT 112 한글 ...
Страница 114: ...1 2 3 4 114 한글 ...
Страница 115: ...5 6 7 8 한글 115 ...
Страница 130: ...1 2 3 4 130 ไทย ...
Страница 131: ...5 6 7 8 ไทย 131 ...
Страница 142: ......
Страница 143: ......